Effect of tissue perfusion on microwave ablation: experimental in vivo study in porcine kidneys.

PURPOSE: To determine the effect of tissue perfusion on microwave ablation lesions in an experimental in vivo study in porcine kidneys.

MATERIALS AND METHODS: Twelve kidneys of six pigs were studied. In each animal, two microwave ablations were created in one kidney without limitation of tissue perfusion (group 1). In the other kidney, two microwave ablations were performed with interruption of blood flow (group 2). All microwave ablations were performed with identical system parameters (eg, temperature control mode, ablation time of 80 s, and temperature of 110°C). The animals were euthanized 3 hours later. The kidneys were harvested and cut into 2-3-mm transverse slices. Microwave ablation zone dimensions (eg, length, width, and volume) and shape (eg, sphericity ratio) and corresponding variability were compared between groups.

RESULTS: Microwave ablation areas were significantly longer (41.6 mm ± 4.0 vs 34.2 mm ± 5.9; P < .01) and wider (16.6 mm ± 1.2 vs 12.2 mm ± 2.1; P < .001) in group 2 than in group 1. Similarly, microwave ablation volume was significantly greater in group 2 compared with group 1 (6.7 cm(3) ± 1.0 vs 3.3 cm(3) ± 1.2; P < .001). Ablation area shapes were similar between groups (sphericity ratio, 2.57 ± 0.42 vs 2.39 ± 0.34). Ablation area variabilities were also comparable between groups (volume variance of 1.32 vs 0.93; sphericity ratio variance of 0.18 vs 0.11).

CONCLUSIONS: After interruption of blood flow, microwave ablation areas are significantly larger than those achieved without limitation of tissue perfusion. Microwave ablation area shape and variability were comparable between study groups.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app