Effect of OCT volume scan density on thickness measurements in diabetic macular edema.

Eye 2011 October
PURPOSE: To evaluate the impact of reducing B-scan frame-sampling density on retinal thickness measurements using spectral domain optical coherence tomography (SD-OCT) in eyes with diabetic macular edema (DME).

METHODS: We retrospectively collected OCT data for 64 eyes of 43 patients undergoing imaging for DME using the Cirrus HD-OCT 512 × 128 macular cube protocol. For each case, raw OCT data were imported into the 3D-OCTOR software, and retinal thickness maps were generated using all 128 B-scans and for lower densities of B-scans ranging from every other scan to only four scans (every 30-s B-scan). Maps were generated before and after manual correction of retinal boundary segmentation errors. The foveal central subfield (FCS) and total macular volume (TMV) values were used to compare thickness maps of varying densities.

RESULTS: The mean difference in FCS retinal thickness and TMV increased as the B-scan density was reduced, particularly when the density was reduced to fewer than 16 B-scans over 6 mm. At a density of 16 B-scans, the mean absolute difference in FCS thickness was 2.43 μm (0.79%), with a maximum of 10.1 μm (4.09%). At this density, the mean difference in TMV was 0.012 mm(3) (0.13%), with a maximum difference of 0.04 mm(3) (0.47%). Manual correction of OCT segmentation errors resulted in a difference in FCS thickness of ≥ 10 μm in only 12.5% of cases, with a maximum difference of 115.7 μm.

CONCLUSION: A minimum of 16 equally spaced B-scans appear necessary to generate retinal thickness measurements similar to those produced using all 128 B-scans in eyes with DME. Manual correction of segmentation errors appeared to have a clinically meaningful effect in a small minority of cases. These results may have implications for the design of SD-OCT imaging and grading protocols in clinical trials of DME, particularly when using multiple SD-OCT instruments that acquire varying numbers of B-scans.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app