Thoracic aorta aneurysm repair using the right axillary approach.

We present a 66-year-old man with a 5.7-cm saccular descending thoracic aortic aneurysm and a smaller 4.6-cm aneurysm just proximal to the celiac artery. The patient was judged to be too risky for open surgical repair because of poor anatomy and health. Previous stenting of the iliac arteries for a kinked aortoiliac open graft precluded conventional endovascular aneurysm repair. The descending thoracic aorta was successfully repaired using endovascular methods with a standard Talent (Medtronic, Los Angeles, Calif) thoracic proximal main stent graft, which was reverse-loaded onto the delivery device and delivered antegrade through the right axillary artery.