Automated attenuation-based tube potential selection for thoracoabdominal computed tomography angiography: improved dose effectiveness.

PURPOSE: To introduce a novel algorithm of automated attenuation-based tube potential selection and to assess its impact on image quality and radiation dose of body computed tomography angiography (CTA).

MATERIALS AND METHODS: In all, 40 patients (mean age 71±11.8 years, body mass index (BMI) 25.7±3.8 kg/m², range 18.8-33.8 kg/m²) underwent 64-slice thoracoabdominal CTA (contrast material: 80 mL, 5 mL/s) using an automated tube potential selection algorithm (CAREkV), which optimizes tube-potential (70-140 kV) and tube-current (138.8±18.6 effective mAs, range 106-177 mAs) based on the attenuation profile of the topogram and on the diagnostic task. Image quality was semiquantitatively assessed by 2 blinded and independent readers (scores 1: excellent to 5: nondiagnostic). Attenuation and noise were measured by another 2 blinded and independent readers. Contrast-to-noise ratio was calculated. The CT dose index (CTDIvol) was recorded and compared with the estimated CTDIvol of a standard 120 kV protocol without using the algorithm in each patient. Selected tube potentials were correlated with BMI and attenuation of the topogram.

RESULTS: Diagnostic image quality was obtained in all patients (excellent: 14; good: 21; moderate: 5; interreader agreement: κ=0.78). Mean attenuation, noise, and contrast-to-noise ratio were 260.8±63.5 Hounsfield units, 15.5±3.3 Hounsfield units, and 14±4.2, respectively, with good to excellent agreement between readers (r=0.50-0.99, P<0.01 each). Automated attenuation-based tube potential selection resulted in a kV-reduction from 120 to 100 kV in 23 patients and to 80 kV in 1 patient, whereas tube potential increased to 140 kV in 1 patient. Automatically selected tube potential showed a significant correlation with both BMI (r=0.427, P<0.05) and attenuation of the topogram (r=0.831, P<0.001). CTDIvol (7.95±2.6 mGy) was significantly lower when using the algorithm compared with the standard 120 kV protocol (10.59±1.8 mGy, P<0.001), corresponding to an overall dose reduction of 25.1%.

CONCLUSION: Automated attenuation-based tube potential selection based on the attenuation profile of the topogram is feasible, provides a diagnostic image quality of body CTA, and reduces overall radiation dose by 25% as compared with a standard protocol with 120 kV.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app