We have located links that may give you full text access.
Second-generation modular acetabular components provide fixation at 10 to 16 years.
Clinical Orthopaedics and related Research 2012 Februrary
BACKGROUND: First-generation modular titanium fiber-metal-coated acetabular components had high rates of wear, pelvic osteolysis, and liner dissociation. Second-generation components were designed to reduce the incidence of these problems but it is unclear whether the changes achieved these goals.
QUESTIONS/PURPOSES: We asked the following questions: (1) Is the risk of revision surgery for loosening, wear, or liner dissociation low with the second-generation acetabular component? (2) Is the rate of pelvic osteolysis low? (3) Can the liner be exchanged without bone cement?
METHODS: We retrospectively reviewed prospectively collected data from 99 patients (118 hips) undergoing THAs with one second-generation modular titanium acetabular component with routine screw fixation and conventional polyethylene. The minimum followup was 10 years (mean, 12 years; range, 10-16 years). We obtained Harris hip scores and examined radiographs for loosening and osteolysis.
RESULTS: At last followup, all acetabular components were well fixed and no titanium shell had been revised or removed. No liner had dissociation. At most recent followup, the mean Harris hip score was 89. We observed pelvic osteolysis in eight hips (7%). There were three reoperations for dislocation (head-liner exchange only) and three loose femoral components revised. Two liners (at 11 and 14 years) were exchanged for wear-pelvic osteolysis.
CONCLUSIONS: This second-generation modular titanium fiber-metal-coated acetabular component with screw fixation had no loosening, no liner dissociation, and a low rate of pelvic osteolysis at 10 to 16 years. Liner exchange is practical without use of cement. We continue to use this component with highly crosslinked polyethylene liners.
LEVEL OF EVIDENCE: Level IV, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.
QUESTIONS/PURPOSES: We asked the following questions: (1) Is the risk of revision surgery for loosening, wear, or liner dissociation low with the second-generation acetabular component? (2) Is the rate of pelvic osteolysis low? (3) Can the liner be exchanged without bone cement?
METHODS: We retrospectively reviewed prospectively collected data from 99 patients (118 hips) undergoing THAs with one second-generation modular titanium acetabular component with routine screw fixation and conventional polyethylene. The minimum followup was 10 years (mean, 12 years; range, 10-16 years). We obtained Harris hip scores and examined radiographs for loosening and osteolysis.
RESULTS: At last followup, all acetabular components were well fixed and no titanium shell had been revised or removed. No liner had dissociation. At most recent followup, the mean Harris hip score was 89. We observed pelvic osteolysis in eight hips (7%). There were three reoperations for dislocation (head-liner exchange only) and three loose femoral components revised. Two liners (at 11 and 14 years) were exchanged for wear-pelvic osteolysis.
CONCLUSIONS: This second-generation modular titanium fiber-metal-coated acetabular component with screw fixation had no loosening, no liner dissociation, and a low rate of pelvic osteolysis at 10 to 16 years. Liner exchange is practical without use of cement. We continue to use this component with highly crosslinked polyethylene liners.
LEVEL OF EVIDENCE: Level IV, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app