Evaluation of hardware-related geometrical distortion in structural MRI at 7 Tesla for image-guided applications in neurosurgery.

RATIONALE AND OBJECTIVES: Geometrical distortion is a well-known problem in structural magnetic resonance imaging (MRI), leading to pixel shifts with variations up to several millimeters. Because the main factors of geometrical distortion are proportional to B(0), MRI spatial encoding distortions tend to increase with higher magnetic field strength. With the increasing prospects of utilizing ultra-high-field MRI (B(0) ≥ 7 Tesla) for neuroimaging and subsequently for image-guided neurosurgical therapy, the evaluation and correction of geometrical distortions occurring in ultra-high-field MRI are essential preconditions for the integration of these data. Hence, we conducted a phantom study to determine hardware-related geometrical distortion in clinically relevant sequences for structural imaging at 7 T MRI and compared the findings to 1.5 T MRI.

MATERIAL AND METHODS: Hardware-related geometrical distortion was evaluated using a MRI phantom (Elekta, Sweden). Both applied scanner systems (Magnetom Avanto 1.5 T and Magnetom 7 T, Siemens Healthcare, Erlangen, Germany) were equipped with similar gradient coils capable of delivering 45 mT/m of maximum amplitude and a slew rate of 220 mT/m/ms. Distortion analysis was performed for various clinically relevant gradient echo and spin echo sequences.

RESULTS: Overall, we found very low mean geometrical distortions at both 7 T and 1.5 T, although single values of up to 1.6 mm were detected. No major differences in mean distortion between the sequences could be found, except significantly higher distortions in turbo spin-echo sequences at 7 T, mainly caused by B(1) inhomogeneities.

CONCLUSION: Hardware-related geometrical distortions at 7 T MRI are relatively small, which may be acceptable for image coregistration or for direct tissue-targeting procedures. Using a subject-specific correction of object-related distortions, an integration of 7 T MRI data into image-guided applications may be feasible.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app