Implementation of a sensitive troponin I assay and risk of recurrent myocardial infarction and death in patients with suspected acute coronary syndrome.

JAMA 2011 March 24
CONTEXT: Although troponin assays have become increasingly more sensitive, it is unclear whether further reductions in the threshold of detection for plasma troponin concentrations will improve clinical outcomes in patients with suspected acute coronary syndrome (ACS).

OBJECTIVE: To determine whether lowering the diagnostic threshold for myocardial infarction (MI) with a sensitive troponin assay could improve clinical outcomes.

DESIGN, SETTING, AND PATIENTS: All consecutive patients admitted with suspected ACS to the Royal Infirmary of Edinburgh, Edinburgh, Scotland, before (n = 1038; February 1-July 31, 2008, during the validation phase) and after (n = 1054; February 1-July 31, 2009, during the implementation phase) lowering the threshold of detection for myocardial necrosis from 0.20 to 0.05 ng/mL with a sensitive troponin I assay were stratified into 3 groups (<0.05 ng/mL, 0.05-0.19 ng/mL, and ≥0.20 ng/mL). During the validation phase, only concentrations above the original diagnostic threshold of 0.20 ng/mL were reported to clinicians.

MAIN OUTCOME MEASURE: Event-free survival (recurrent MI and death) at 1 year in patients grouped by plasma troponin concentrations.

RESULTS: Plasma troponin concentrations were less than 0.05 ng/mL in 1340 patients (64%), 0.05 to 0.19 ng/mL in 170 patients (8%), and 0.20 ng/mL or more in 582 patients (28%). During the validation phase, 39% of patients with plasma troponin concentrations of 0.05 to 0.19 ng/mL were dead or had recurrent MI at 1 year compared with 7% and 24% of those patients with troponin concentrations of less than 0.05 ng/mL (P < .001) or 0.20 ng/mL or more (P = .007), respectively. During the implementation phase, lowering the diagnostic threshold to 0.05 ng/mL was associated with a lower risk of death and recurrent MI (from 39% to 21%) in patients with troponin concentrations of 0.05 to 0.19 ng/mL (odds ratio, 0.42; 95% confidence interval, 0.24-0.84; P = .01).

CONCLUSIONS: In patients with suspected ACS, implementation of a sensitive troponin assay increased the diagnosis of MI and identified patients at high risk of recurrent MI and death. Lowering the diagnostic threshold of plasma troponin was associated with major reductions in morbidity and mortality.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app