EVALUATION STUDIES
JOURNAL ARTICLE
Add like
Add dislike
Add to saved papers

Radiographic analysis of a hand-held surgical navigation system for tibial resection in total knee arthroplasty.

Tibial intramedullary or extramedullary alignment guides have not been shown to be highly accurate in performing the tibial resection in total knee arthroplasty (TKA). Since May 2010, a total of 42 knees underwent a TKA using a hand-held, accelerometer-based surgical navigation system for performing the tibial resection (KneeAlign; OrthAlign Inc, Aliso Viejo, Calif). Postoperative standing anteroposterior hip-to-ankle and lateral knee-to-ankle radiographs demonstrated that 97.6% of the tibial components were placed within 90° ± 2° to the mechanical axis in the coronal plane, and 96.2% of the components were placed within 3° ± 2° to the mechanical axis in the sagittal plane. The KneeAlign greatly improves the accuracy of tibial component alignment in TKA.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app