Diagnostic accuracy of point-of-care testing for diabetic ketoacidosis at emergency-department triage: {beta}-hydroxybutyrate versus the urine dipstick.
Diabetes Care 2011 April
OBJECTIVE: In the emergency department, hyperglycemic patients are screened for diabetic ketoacidosis (DKA) via a urine dipstick. In this prospective study, we compared the test characteristics of point-of-care β-hydroxybutyrate (β-OHB) analysis with the urine dipstick.
RESEARCH DESIGN AND METHODS: Emergency-department patients with blood glucose ≥250 mg/dL had urine dipstick, chemistry panel, venous blood gas, and capillary β-OHB measurements. DKA was diagnosed according to American Diabetes Association criteria.
RESULTS: Of 516 hyperglycemic subjects, 54 had DKA. The urine dipstick had a sensitivity of 98.1% (95% CI 90.1-100), a specificity of 35.1% (30.7-39.6), a positive predictive value of 15% (11.5-19.2), and a negative predictive value of 99.4% (96.6-100) for DKA. Using the manufacturer-suggested cutoff of >1.5 mmol/L, β-OHB had a sensitivity of 98.1% (90.1-100), a specificity of 78.6% (74.5-82.2), a positive predictive value of 34.9% (27.3-43), and a negative predictive value of 99.7% (98.5-100) for DKA.
CONCLUSIONS: Point-of-care β-OHB and the urine dipstick are equally sensitive for detecting DKA (98.1%). However, β-OHB is more specific (78.6 vs. 35.1%), offering the potential to significantly reduce unnecessary DKA work-ups among hyperglycemic patients in the emergency department.
RESEARCH DESIGN AND METHODS: Emergency-department patients with blood glucose ≥250 mg/dL had urine dipstick, chemistry panel, venous blood gas, and capillary β-OHB measurements. DKA was diagnosed according to American Diabetes Association criteria.
RESULTS: Of 516 hyperglycemic subjects, 54 had DKA. The urine dipstick had a sensitivity of 98.1% (95% CI 90.1-100), a specificity of 35.1% (30.7-39.6), a positive predictive value of 15% (11.5-19.2), and a negative predictive value of 99.4% (96.6-100) for DKA. Using the manufacturer-suggested cutoff of >1.5 mmol/L, β-OHB had a sensitivity of 98.1% (90.1-100), a specificity of 78.6% (74.5-82.2), a positive predictive value of 34.9% (27.3-43), and a negative predictive value of 99.7% (98.5-100) for DKA.
CONCLUSIONS: Point-of-care β-OHB and the urine dipstick are equally sensitive for detecting DKA (98.1%). However, β-OHB is more specific (78.6 vs. 35.1%), offering the potential to significantly reduce unnecessary DKA work-ups among hyperglycemic patients in the emergency department.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app