Evaluation of the Bullard, GlideScope, Viewmax, and Macintosh laryngoscopes using a cadaver model to simulate the difficult airway.

STUDY OBJECTIVE: To assess the performance and cervical (C)-spine movement associated with laryngoscopy using the Bullard laryngoscope (BL), GlideScope videolaryngoscope (GVL), Viewmax, and Macintosh laryngoscopes during conditions of a) unrestricted and b) restricted C-spine and temporomandibular joint (TMJ) mobility.

DESIGN: Prospective, controlled, randomized, crossover study.

SETTING: University teaching hospital.

SUBJECTS: 21 cadavers with intact C-spine anatomy.

INTERVENTIONS: Each cadaver underwent to total of 8 intubation attempts to complete the intubation protocol using all four devices under unrestricted and restricted C-spine and TMJ mobility.

MEASUREMENTS: Laryngoscopic view was graded using the modified Cormack-Lehane system. Time to best laryngoscopic view and total time to intubation were recorded. C-spine movement was measured between McGregor's line and each vertebra from radiographs taken at baseline and at best laryngoscopic view.

MAIN RESULTS: During both intubating conditions, the BL achieved the highest number of modified Cormack-Lehane grade 1 and 2A laryngoscopic views as compared to the other three devices (P < 0.05) and had fewer intubation failures than the Viewmax or Macintosh laryngoscopes (P < 0.05). The GVL had superior laryngoscopic performance as compared to the Viewmax and Macintosh laryngoscopes (P < 0.05) and had fewer intubation failures than those two devices (P < 0.05). All devices except the Macintosh laryngoscope in restricted mobility achieved median times to intubation in less than 30 seconds. For both conditions, BL showed the least total absolute movement between Occiput/C1 and C3/C4 of all the devices (all P < 0.05). Most of the difference was seen at C1/C2.

CONCLUSIONS: In cadavers with unrestricted and restricted C-spine mobility, the BL provided superior laryngoscopic views, comparable intubating times, and less C-spine movement than the GVL, Viewmax, or Macintosh laryngoscopes.

Full text links

For the best experience, use the Read mobile app

Group 7SearchHeart failure treatmentPapersTopicsCollectionsEffects of Sodium-Glucose Cotransporter 2 Inhibitors for the Treatment of Patients With Heart Failure Importance: Only 1 class of glucose-lowering agents-sodium-glucose cotransporter 2 (SGLT2) inhibitors-has been reported to decrease the risk of cardiovascular events primarily by reducingSeptember 1, 2017: JAMA CardiologyAssociations of albuminuria in patients with chronic heart failure: findings in the ALiskiren Observation of heart Failure Treatment study.CONCLUSIONS: Increased UACR is common in patients with heart failure, including non-diabetics. Urinary albumin creatininineJul, 2011: European Journal of Heart FailureRandomized Controlled TrialEffects of Liraglutide on Clinical Stability Among Patients With Advanced Heart Failure and Reduced Ejection Fraction: A Randomized Clinical Trial.Review

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app