BACKGROUND: ApneaLink (AL) is a single-channel type-4 device that measures airflow. A limited number of studies have assessed AL's usefulness in diagnosing obstructive sleep apnea (OSA) using automated scoring alone. This study was conducted to assess the sensitivity and specificity of AL in a selected group of people with clinical suspicion of OSA, using both automatic and manual scoring and comparing the results with those obtained for polysomnography (PSG).

MATERIAL/METHODS: Simultaneous overnight recordings of in-laboratory PSG and AL were performed for 95 patients (mean age 46.3 ± 12.6 yr) with a high clinical suspicion of OSA. PSG was scored manually according to the American Academy of Sleep Medicine (AASM) guidelines. AL data were analyzed automatically using a manufacturer-provided computerized algorithm. Manual scoring of the AL flow signal followed the AASM guidelines for reduction in flow.

RESULTS: The mean apnea hypopneas index (AHI) for PSG was 34.1 ± 32.4/hr. The AL Auto-AHI was 20.1 ± 25.2/h, and the AL Manual-AHI was 39.5 ± 30.4/h. The Pearson correlation coefficients were r=0.883 between PSG-AHI and AL Auto-AHI, and r=0.966 between PSG-AHI and AL Manual-AHI. At AHIs of 5, 10, 15, and 30, the AL Auto sensitivity/specificity was 0.79/0.68, 0.70/0.89, 0.64/0.94 and 0.63/0.98, and the AL Manual sensitivity/specificity was 1.00/0.43, 1.00/0.56, 0.98/0.58 and 1.00/0.80.

CONCLUSIONS: Combining auto and manual scoring of data (automatic scoring followed by manual scoring) recorded by single-channel ApneaLink provides good diagnostic agreement with conventional PSG recordings.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
21278698
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"