Test-retest reliability of the GAITRite system in people with stroke undergoing rehabilitation.
PURPOSE: To determine the test-retest reliability of the GAITRite system in measuring spatio-temporal gait parameters in people with stroke undergoing inpatient rehabilitation and to investigate whether walking ability influences test-retest reliability.
METHOD: A within-subject repeated measures study was conducted. Twenty-one individuals with first stroke undergoing inpatient rehabilitation participated. The Motor Assessment Scale Item 5 score was used to categorise participants into poorer (score 3 or 4) or better walkers (score 5 or 6). Participants made two passes at a self-selected gait speed over the GAITRite mat on two separate occasions. Intra-class correlation coefficients (ICC(2,1)) and 95% limits of agreement for spatio-temporal gait parameters between the two occasions and groups were determined.
RESULTS: No differences in gait parameters between the two sessions (p > 0.15) were found. Test-retest reliability for the whole group was good to excellent, ICC(2,1) ranging from 0.72 to 0.94. Walking ability did influence these findings with fair to good reliability demonstrated for both poorer (ICC(2,1) > 0.57) and better walkers (ICC(2,1) > 0.41).
CONCLUSIONS: It appears that the GAITRite has good test-retest reliability in this group of people with first stroke undergoing rehabilitation. Test reliability seems to be slightly higher for poorer walkers than better walkers.
METHOD: A within-subject repeated measures study was conducted. Twenty-one individuals with first stroke undergoing inpatient rehabilitation participated. The Motor Assessment Scale Item 5 score was used to categorise participants into poorer (score 3 or 4) or better walkers (score 5 or 6). Participants made two passes at a self-selected gait speed over the GAITRite mat on two separate occasions. Intra-class correlation coefficients (ICC(2,1)) and 95% limits of agreement for spatio-temporal gait parameters between the two occasions and groups were determined.
RESULTS: No differences in gait parameters between the two sessions (p > 0.15) were found. Test-retest reliability for the whole group was good to excellent, ICC(2,1) ranging from 0.72 to 0.94. Walking ability did influence these findings with fair to good reliability demonstrated for both poorer (ICC(2,1) > 0.57) and better walkers (ICC(2,1) > 0.41).
CONCLUSIONS: It appears that the GAITRite has good test-retest reliability in this group of people with first stroke undergoing rehabilitation. Test reliability seems to be slightly higher for poorer walkers than better walkers.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app