Elastomer femoroplasty prevents hip fracture displacement In vitro biomechanical study comparing two minimal invasive femoroplasty techniques.
Clinical Biomechanics 2011 June
UNLABELLED: The purpose of this study was to test femur strength and the ability to prevent fracture displacement of two minimal invasive Elastomer femoroplasty techniques.
METHODS: A total of sixteen fixed human cadaveric femur pairs were used. From each pair one femur was randomly assigned for Elastomer femoroplasty. In these femora we drilled a 3.5mm entrance in the lateral cortex. Cavities for the Elastomer were created by: group A, balloon and group B an excentric drill. All femora were fractured by simulating a fall on the greater trochanter. Neck-shaft-angles on plain anterior posterior radiographs were measured to determine fracture displacement.
FINDINGS: There was no significant difference in fracture load between controls and treated femora for group A, 2904N (SD 1091) versus 2803N (SD 627) and group B, 2773N (SD 747) versus 2597N (SD 834). In group A the mean displacement was 35° (SD 14) for the control femora and 3° (SD 2) for the treated femora (P<0.001). In group B the mean displacement was 38° (SD 10) for the controls and 8° (SD 13) for the treated femora (P<0.001).
INTERPRETATION: The results of this study show that minimal invasive Elastomer femoroplasty prevents fracture displacement of the proximal femur. We found no significant compromise in load-to-fracture after minimal invasive balloon or excentric drill femoroplasty.
METHODS: A total of sixteen fixed human cadaveric femur pairs were used. From each pair one femur was randomly assigned for Elastomer femoroplasty. In these femora we drilled a 3.5mm entrance in the lateral cortex. Cavities for the Elastomer were created by: group A, balloon and group B an excentric drill. All femora were fractured by simulating a fall on the greater trochanter. Neck-shaft-angles on plain anterior posterior radiographs were measured to determine fracture displacement.
FINDINGS: There was no significant difference in fracture load between controls and treated femora for group A, 2904N (SD 1091) versus 2803N (SD 627) and group B, 2773N (SD 747) versus 2597N (SD 834). In group A the mean displacement was 35° (SD 14) for the control femora and 3° (SD 2) for the treated femora (P<0.001). In group B the mean displacement was 38° (SD 10) for the controls and 8° (SD 13) for the treated femora (P<0.001).
INTERPRETATION: The results of this study show that minimal invasive Elastomer femoroplasty prevents fracture displacement of the proximal femur. We found no significant compromise in load-to-fracture after minimal invasive balloon or excentric drill femoroplasty.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app