Lack of agreement between pulmonary arterial thermodilution cardiac output and the pressure recording analytical method in postoperative cardiac surgery patients.

BACKGROUND: Pulse-contour analysis method (PCM) cardiac output (CO) monitors are increasingly used for CO monitoring during anaesthesia and in the critically ill. Very recently, several systems have been introduced that do not need calibration; among them the pressure recording analytical method (PRAM). Sparse data comparing the accuracy of the PRAM-CO with conventional thermodilution CO (ThD-CO) in cardiac surgery patients are available.

METHODS: In this prospective comparison study, paired CO measurements with a pulmonary artery catheter and a PRAM monitoring set were obtained 20-30 min apart (t1, t2) in 23 extubated patients on the first postoperative day after cardiac surgery. Data were analysed by the Bland-Altman method.

RESULTS: A total of 46 paired CO measurements (23 for each interval) were collected. The Bland-Altman analysis showed a mean difference (bias) of 0.0 litre min(-1) and limits of agreement (1.96 sd) of 4.53 to -4.54 litre min(-1) [upper 95% confidence interval (CI), 3.26-5.80; lower 95% CI, -5.8 to -3.27]. The percentage error (1.96 sd/mean of the reference method) was 87%.

CONCLUSIONS: These results question the reliability of the PRAM technology for the determination of CO in postoperative cardiac surgery patients.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app