Flexible pressure delivery modification of continuous positive airway pressure for obstructive sleep apnea does not improve compliance with therapy: systematic review and meta-analysis.
Chest 2011 June
BACKGROUND: Continuous positive airway pressure (CPAP) is the first-line therapy for obstructive sleep apnea (OSA), but patient compliance is a major barrier to long-term effectiveness. Flexible pressure delivery of PAP reduces pressure during early exhalation with the aim of improving comfort and, therefore, compliance, leading to subsequent symptoms improvement.
METHODS: We undertook a systematic literature search of PubMed (January 1, 2000, to July 11, 2010) to identify all randomized trials comparing flexible and standard CPAP in adult patients with OSA with at least 1-week follow-up. Either we or the original trial investigators extracted means, SEs, and sample sizes for all relevant outcome measures. We then performed meta-analyses quantifying improvements in objective compliance and symptoms as measured by the Epworth Sleepiness Scale (ESS), the Maintenance of Wakefulness Test (MWT), and the Psychomotor Vigilance Task (PVT).
RESULTS: We found 10 relevant trials (599 patients). Meta-analysis of the seven trials where we could extract compliance data (514 patients) indicated that flexible pressure did not improve compliance compared with CPAP in either the parallel (0.16 h; 95% CI, -0.09-0.42; P = .21) or the crossover trials (0.20 h; 95% CI, -0.26-0.66; P = .39). Flexible pressure caused no improvement over CPAP in any secondary outcome (ESS, MWT, PVT, and residual OSA, all P > .05).
CONCLUSIONS: Flexible pressure modification neither significantly improves compliance with CPAP in patients with OSA nor significantly improves patient outcomes beyond the effects of CPAP. Unfortunately, we were unable to locate compliance data in the correct format for three out of the 10 suitable trials.
METHODS: We undertook a systematic literature search of PubMed (January 1, 2000, to July 11, 2010) to identify all randomized trials comparing flexible and standard CPAP in adult patients with OSA with at least 1-week follow-up. Either we or the original trial investigators extracted means, SEs, and sample sizes for all relevant outcome measures. We then performed meta-analyses quantifying improvements in objective compliance and symptoms as measured by the Epworth Sleepiness Scale (ESS), the Maintenance of Wakefulness Test (MWT), and the Psychomotor Vigilance Task (PVT).
RESULTS: We found 10 relevant trials (599 patients). Meta-analysis of the seven trials where we could extract compliance data (514 patients) indicated that flexible pressure did not improve compliance compared with CPAP in either the parallel (0.16 h; 95% CI, -0.09-0.42; P = .21) or the crossover trials (0.20 h; 95% CI, -0.26-0.66; P = .39). Flexible pressure caused no improvement over CPAP in any secondary outcome (ESS, MWT, PVT, and residual OSA, all P > .05).
CONCLUSIONS: Flexible pressure modification neither significantly improves compliance with CPAP in patients with OSA nor significantly improves patient outcomes beyond the effects of CPAP. Unfortunately, we were unable to locate compliance data in the correct format for three out of the 10 suitable trials.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
Read by QxMD is copyright © 2021 QxMD Software Inc. All rights reserved. By using this service, you agree to our terms of use and privacy policy.
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app