Evaluation Studies
Journal Article
Multicenter Study
Research Support, U.S. Gov't, P.H.S.
Add like
Add dislike
Add to saved papers

Bridging children of all sizes to cardiac transplantation: the initial multicenter North American experience with the Berlin Heart EXCOR ventricular assist device.

BACKGROUND: Beginning in 2000 and accelerating in 2004, the Berlin Heart EXCOR (Berlin Heart Inc Woodlands, TX) became the first pediatric-specific ventricular assist device (VAD) applied throughout North America for children of all sizes. This retrospective study analyzed the initial Berlin Heart EXCOR pediatric experience as a bridge to transplantation.

METHODS: Between June 2000 and May 2007, 97 EXCOR VADs were implanted in North America at 29 different institutions. The analysis is limited to 73 patients (75%) from 17 institutions, for which retrospective data were available.

RESULTS: Median age and weight at VAD implant were 2.1 years (range, 12 days-17.8 years) and 11 kg (range, 3-87.6 kg), respectively. The primary diagnoses were dilated cardiomyopathy in 42 (58%), congenital heart disease in 19 (26%), myocarditis in 7 (10%), and other cardiomyopathies in 5 (7%). Pre-implant clinical condition was critical cardiogenic shock in 38 (52%), progressive decline in 33 (45%), or other in 2 (3%). Extracorporeal membrane oxygenation was used as a bridge to EXCOR in 22 patients (30%). Device selection was left VAD (LVAD) in 42 (57%) and biventricular assist devices (BiVAD) in 31 (43%). The EXCOR bridged 51 patients (70%) to transplant and 5 (7%) to recovery. Mortality on the EXCOR was 23% (n = 17) overall, including 35% (11 of 31) in BiVAD vs 14% (6 of 42) in LVAD patients (p = 0.003). Multivariate analysis showed younger age and BiVAD support were significant risk factors for death while on the EXCOR.

CONCLUSIONS: This limited but large preliminary North American experience with the Berlin Heart EXCOR VAD as a bridge to cardiac transplantation for children of all ages and sizes points to the feasibility of this approach. The prospective investigational device evaluation trial presently underway will further characterize the safety and efficacy of the EXCOR as a bridge to pediatric cardiac transplantation.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app