Myocardial rotation but not circumferential strain is transducer angle dependent: a speckle tracking echocardiography study.
Echocardiography 2010 August
BACKGROUND: Doppler derived strain analysis has been shown to be angle dependent. Speckle tracking analysis using 2D echocardiographic images is thought to provide angle independent parameters of regional and global myocardial function. This study sought to evaluate whether myocardial circumferential strain and rotation derived from automatic frame-by-frame tracking of natural acoustic markers is dependent on angulation of the transducer.
METHODS: In 48 healthy volunteers (mean age 36 ± 3 years, 20 male) parasternal short-axis views at apical level were obtained as follows: at the standard parasternal position (5th intercostal space) with a most possible circular short-axis image of the left ventricle (angulation 1), at an angulation of the transducer by 20° from this standard position to the apex (angulation 2) and at an angulation of the transducer by 20° to the base of the left ventricle (angulation 3). Using an automatic frame-by-frame tracking system of natural acoustic echocardiographic markers (EchoPAC, GE Ultrasound, Horton, Norway) circumferential strain and rotation were calculated for six segments within a short-axis circumference.
RESULTS: Image quality was sufficient for acquisition and analysis of images at all three-transducer angulation in 90% of the analyzed segments. Rotation was measured to be 7.7 ± 1.2° at angulation 1, 2.7 ± 0.9° at angulation 2 and 4.3 ± 1.1° at angulation 3 (p < 0.05). Average circumferential strain data was found to be -27.2 ± 5.1% at angulation 1, -26.5 ± 3.8% at angulation 2 and 28.9 ± 4.4% at angulation 3 (p = 0.287).
CONCLUSION: Circumferential strain analysis is not dependent on transducer angulation. In contrast, determination of myocardial rotation is dependent on transducer angulation. Therefore, accurate transducer angulation has to be taken care of if rotation measurements are performed.
METHODS: In 48 healthy volunteers (mean age 36 ± 3 years, 20 male) parasternal short-axis views at apical level were obtained as follows: at the standard parasternal position (5th intercostal space) with a most possible circular short-axis image of the left ventricle (angulation 1), at an angulation of the transducer by 20° from this standard position to the apex (angulation 2) and at an angulation of the transducer by 20° to the base of the left ventricle (angulation 3). Using an automatic frame-by-frame tracking system of natural acoustic echocardiographic markers (EchoPAC, GE Ultrasound, Horton, Norway) circumferential strain and rotation were calculated for six segments within a short-axis circumference.
RESULTS: Image quality was sufficient for acquisition and analysis of images at all three-transducer angulation in 90% of the analyzed segments. Rotation was measured to be 7.7 ± 1.2° at angulation 1, 2.7 ± 0.9° at angulation 2 and 4.3 ± 1.1° at angulation 3 (p < 0.05). Average circumferential strain data was found to be -27.2 ± 5.1% at angulation 1, -26.5 ± 3.8% at angulation 2 and 28.9 ± 4.4% at angulation 3 (p = 0.287).
CONCLUSION: Circumferential strain analysis is not dependent on transducer angulation. In contrast, determination of myocardial rotation is dependent on transducer angulation. Therefore, accurate transducer angulation has to be taken care of if rotation measurements are performed.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app