COMPARATIVE STUDY
JOURNAL ARTICLE
TECHNICAL REPORT
Hand-carried ultrasound devices in pediatric cardiology: clinical experience with three different devices in 110 patients.
Journal of the American Society of Echocardiography 2010 December
BACKGROUND: The aims of this study were to determine the usefulness of hand-carried ultrasound devices in pediatric cardiology and to compare the performance of three different hand-carried ultrasound devices in a pediatric cardiology outpatient clinic and intensive care unit.
METHODS: One hundred ten patients (49 male; mean age, 6.4 ± 5.2 years; range 0.1-38 years) with congenital heart defects or innocent heart murmurs were examined using Siemens Acuson P10, Siemens Acuson P50, and Philips CX 50 systems. The quality of images and the accuracy of B-mode measurements were compared with those obtained using a standard echocardiographic system (Philips iE33).
RESULTS: Fifty-nine patients were examined with the Siemens Acuson P10, 29 with the Siemens Acuson P50, and 22 with the Philips CX 50 system. There were no significant differences in B-mode measurements. The Acuson P10 system, however, showed significantly lower image quality, with 64.54% of all studies considered of excellent quality compared with 92.83% with the Acuson P50 and 95.52% with the CX 50 (P < .05) and a mean quality score (1 = fair, 5 = excellent) of 3.5 versus 4.57 with the Acuson P50 and 4.86 with the CX 50 (P < .05). This was attributed to the limited capacity for accurate diagnosis in children with body weights < 10 kg and complex heart defects.
CONCLUSION: Hand-carried ultrasound devices represent a valuable alternative to standard echocardiographic systems in pediatric cardiology. In particular, systems including all echocardiographic modalities offer unlimited versatility in outpatient and intensive care.
METHODS: One hundred ten patients (49 male; mean age, 6.4 ± 5.2 years; range 0.1-38 years) with congenital heart defects or innocent heart murmurs were examined using Siemens Acuson P10, Siemens Acuson P50, and Philips CX 50 systems. The quality of images and the accuracy of B-mode measurements were compared with those obtained using a standard echocardiographic system (Philips iE33).
RESULTS: Fifty-nine patients were examined with the Siemens Acuson P10, 29 with the Siemens Acuson P50, and 22 with the Philips CX 50 system. There were no significant differences in B-mode measurements. The Acuson P10 system, however, showed significantly lower image quality, with 64.54% of all studies considered of excellent quality compared with 92.83% with the Acuson P50 and 95.52% with the CX 50 (P < .05) and a mean quality score (1 = fair, 5 = excellent) of 3.5 versus 4.57 with the Acuson P50 and 4.86 with the CX 50 (P < .05). This was attributed to the limited capacity for accurate diagnosis in children with body weights < 10 kg and complex heart defects.
CONCLUSION: Hand-carried ultrasound devices represent a valuable alternative to standard echocardiographic systems in pediatric cardiology. In particular, systems including all echocardiographic modalities offer unlimited versatility in outpatient and intensive care.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app