Stabilization of the posteromedial fragment in bicondylar tibial plateau fractures: a mechanical comparison of locking and nonlocking single and dual plating methods.
Journal of Trauma 2010 July
BACKGROUND: The objective of this study is to compare locking and nonlocking single and dual plating constructs in maintaining posteromedial fragment reduction in a bicondylar tibial plateau fracture model. We hypothesized that posteromedial fragment fixation with medial and lateral nonlocked constructs would tolerate higher loads than with lateral locked constructs alone.
METHODS: Thirty composite tibiae were fractured (AO 41-C1.3). Six constructs were tested: (1) lateral 3.5-mm conventional nonlocking proximal tibial plate (CP); (2) CP + posteromedial 3.5-mm limited contact dynamic compression plate; (3) CP + posteromedial 1/3 tubular plate (CP + 1/3 tubular); (4) 3.5-mm Zimmer Proximal Tibial Locking plate; (5) 3.5-mm Synthes Locking Compression plate; and (6) Less Invasive Stabilization System tibial plate. Specimens were cyclically loaded to failure or a maximum of 4000 N.
RESULTS: Failure occurred at the posteromedial fragment first. The CP + 1/3 tubular had the highest average load to failure (3040 N). The CP + 1/3 tubular demonstrated higher load at failure compared with the 3.5-mm Synthes Locking Compression plate (p = 0.0060) and the Less Invasive Stabilization System (p = 0.0360). The CP + 1/3 tubular did not demonstrate a difference in load at failure when compared with the CP (p = 0.4225), the CP + posteromedial 3.5-mm limited contact dynamic compression plate (p = 0.4799), or the 3.5-mm Zimmer Proximal Tibial Locking plate (p = 0.1595).
CONCLUSIONS: The posteromedial fragment tolerated higher loads with the CP + 1/3 tubular plate construct. The superiority of this construct may be caused by unreliable penetration of this fragment by the lateral locking screws.
METHODS: Thirty composite tibiae were fractured (AO 41-C1.3). Six constructs were tested: (1) lateral 3.5-mm conventional nonlocking proximal tibial plate (CP); (2) CP + posteromedial 3.5-mm limited contact dynamic compression plate; (3) CP + posteromedial 1/3 tubular plate (CP + 1/3 tubular); (4) 3.5-mm Zimmer Proximal Tibial Locking plate; (5) 3.5-mm Synthes Locking Compression plate; and (6) Less Invasive Stabilization System tibial plate. Specimens were cyclically loaded to failure or a maximum of 4000 N.
RESULTS: Failure occurred at the posteromedial fragment first. The CP + 1/3 tubular had the highest average load to failure (3040 N). The CP + 1/3 tubular demonstrated higher load at failure compared with the 3.5-mm Synthes Locking Compression plate (p = 0.0060) and the Less Invasive Stabilization System (p = 0.0360). The CP + 1/3 tubular did not demonstrate a difference in load at failure when compared with the CP (p = 0.4225), the CP + posteromedial 3.5-mm limited contact dynamic compression plate (p = 0.4799), or the 3.5-mm Zimmer Proximal Tibial Locking plate (p = 0.1595).
CONCLUSIONS: The posteromedial fragment tolerated higher loads with the CP + 1/3 tubular plate construct. The superiority of this construct may be caused by unreliable penetration of this fragment by the lateral locking screws.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app