COMPARATIVE STUDY
JOURNAL ARTICLE

A comparison of in-room computerized tomography options for detection of fiducial markers in prostate cancer radiotherapy

Rebecca Owen, Farshad Foroudi, Tomas Kron, Alvin Milner, Jennifer Cox, Jim Cramb, Li Zhu, Gillian Duchesne
International Journal of Radiation Oncology, Biology, Physics 2010 July 15, 77 (4): 1248-56
20610045

PURPOSE: To compare volumetric in-room computed tomography (CT) and kilovoltage (kV) cone-beam CT (CBCT) to planar imaging with respect to their ability to localize fiducial markers (FMs) for radiotherapy of prostate cancer.

METHODS AND MATERIALS: Image guidance options from two linear accelerators were compared in terms of identifying the center of gravity (COG) of FMs from the isocenter: a Siemens Primatom, where the couch is rotated 180 degrees from the treatment isocenter to the in-room CT vs. electronic portal imaging (EPI); and a Varian OBI system, where kV CBCT, EPI, and planar kV radiographs were compared. In all, 387 image pairs (CBCT = 133; CT = 254) from 18 patients were analyzed. A clinical tolerance of 3 mm was predefined as the acceptable threshold for agreement.

RESULTS: COG location on in-room CT and EPI was in agreement 96.9%, 85.8%, and 89.0% of the time in the left-right (LR), superior-inferior (SI), and anterior-posterior (AP) directions, respectively, vs. 99.2%, 91.7%, and 93.2% for the CBCT and EPI analysis. The CBCT vs. kV radiographs were in agreement 100% (LR), 85.4% (SI), and 88.5% (AP), and EPI vs. kV radiographs were in agreement 100% (LR), 94.6% (SI), and 91.5% (AP) of the time.

CONCLUSION: Identification of FMs on volumetric or planar images was found to be not equivalent (+/-3 mm) using either linear accelerator. Intrafraction prostate motion, interpretation of FM location, and spatial properties of images are contributing factors. Although in-room CT has superior image quality, the process of realigning the treatment couch to acquire a CT introduces an error, highlighting the benefits of a single isocentric system.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
20610045
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"