Improved left ventricular unloading and circulatory support with synchronized pulsatile left ventricular assistance compared with continuous-flow left ventricular assistance in an acute porcine left ventricular failure model.

OBJECTIVE: Controversy exists regarding the optimal pumping method for left ventricular assist devices. The purpose of this investigation was to test the hypothesis that pulsatile left ventricular assist synchronized to the cardiac cycle provides superior left ventricular unloading and circulatory support compared with continuous-flow left ventricular assist devices at the same level of ventricular assist device flow.

METHODS: Seven male pigs were used to evaluate left ventricular assist device function using the TORVAD synchronized pulsatile-flow pump (Windmill Cardiovascular Systems, Inc, Austin, Tex) compared with the Bio-Medicus BPX-80 continuous-flow centrifugal pump (Medtronic, Inc, Minneapolis, Minn). Experiments were carried out under general anesthesia, and animals were instrumented via a median sternotomy. Hemodynamic measurements were obtained in the control state and with left ventricular assistance using the TORVAD and BPX-80 individually. Left ventricular failure was induced with suture ligation of the mid-left anterior descending coronary artery, and hemodynamic measurements were repeated.

RESULTS: During left ventricular assist device support, mean aortic pressure and total cardiac output were higher and left atrial pressure was lower with pulsatile compared with continuous flow at the same ventricular assist device flow rate. During ischemic left ventricular failure, pulsatile left ventricular support resulted in higher total cardiac output (5.58 ± 1.58 vs 5.12 ± 1.19, P < .05), higher mean aortic pressure (67.8 ± 14 vs 60.2 ± 10, P < .05), and lower left atrial pressure (11.5 ± 3.5 vs 13.9 ± 6.0, P < .05) compared with continuous flow at the same left ventricular assist device flow rate.

CONCLUSION: Synchronized, pulsatile left ventricular assistance produces superior left ventricular unloading and circulatory support compared with continuous-flow left ventricular assist at the same flow rates.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app