Assessment of planning target volume margins for intensity-modulated radiotherapy of the prostate gland: role of daily inter- and intrafraction motion.

PURPOSE: To determine planning target volume margins for prostate intensity-modulated radiotherapy based on inter- and intrafraction motion using four daily localization techniques: three-point skin mark alignment, volumetric imaging with bony landmark registration, volumetric imaging with implanted fiducial marker registration, and implanted electromagnetic transponders (beacons) detection.

METHODS AND MATERIALS: Fourteen patients who underwent definitive intensity-modulated radiotherapy for prostate cancer formed the basis of this study. Each patient was implanted with three electromagnetic transponders and underwent a course of 39 treatment fractions. Daily localization was based on three-point skin mark alignment followed by transponder detection and patient repositioning. Transponder positioning was verified by volumetric imaging with cone-beam computed tomography of the pelvis. Relative motion between the prostate gland and bony anatomy was quantified by offline analyses of daily cone-beam computed tomography. Intratreatment organ motion was monitored continuously by the Calypso® System for quantification of intrafraction setup error.

RESULTS: As expected, setup error (that is, inter- plus intrafraction motion, unless otherwise stated) was largest with skin mark alignment, requiring margins of 7.5 mm, 11.4 mm, and 16.3 mm, in the lateral (LR), longitudinal (SI), and vertical (AP) directions, respectively. Margin requirements accounting for intrafraction motion were smallest for transponder detection localization techniques, requiring margins of 1.4 mm (LR), 2.6 mm (SI), and 2.3 mm (AP). Bony anatomy alignment required 2.1 mm (LR), 9.4 mm (SI), and 10.5 mm (AP), whereas image-guided marker alignment required 2.8 mm (LR), 3.7 mm (SI), and 3.2 mm (AP). No marker migration was observed in the cohort.

CONCLUSION: Clinically feasible, rapid, and reliable tools such as the electromagnetic transponder detection system for pretreatment target localization and, subsequently, intratreatment target location monitoring allow clinicians to reduce irradiated volumes and facilitate safe dose escalation, where appropriate.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app