Whole-heart coronary magnetic resonance angiography at 3.0T using short-TR steady-state free precession, vastly undersampled isotropic projection reconstruction

Jingsi Xie, Peng Lai, Himanshu Bhat, Debiao Li
Journal of Magnetic Resonance Imaging: JMRI 2010, 31 (5): 1230-5
20432361

PURPOSE: To evaluate the feasibility of improving 3.0T steady-state free precession (SSFP) whole-heart coronary magnetic resonance angiography (MRA) using short-TR (repetition time) VIPR (vastly undersampled isotropic projection reconstruction).

MATERIALS AND METHODS: SSFP is highly sensitive to field inhomogeneity. VIPR imaging uses nonselective radiofrequency pulses, allowing short TR and reduced banding artifacts, while achieving isotropic 3D resolution. Coronary artery imaging was performed in nine healthy volunteers using SSFP VIPR. TR was reduced to 3.0 msec with an isotropic spatial resolution of 1.3 x 1.3 x 1.3 mm(3). Image quality, vessel sharpness, and lengths of major coronary arteries were measured. Comparison between SSFP using Cartesian trajectory and SSFP using VIPR trajectory was performed in all volunteers.

RESULTS: Short-TR SSFP VIPR resulted in whole-heart images without any banding artifacts, leading to excellent coronary artery visualization. The average image quality score for VIPR-SSFP was 3.12 +/- 0.42 out of four while that for Cartesian SSFP was 0.92 +/- 0.61. A significant improvement (P < 0.05) in image quality was shown by Wilcoxon comparison. The visualized coronary artery lengths for VIPR-SSFP were: 10.13 +/- 0.79 cm for the left anterior descending artery (LAD), 7.90 +/- 0.91 cm for the left circumflex artery (LCX), 7.50 +/- 1.65 cm for the right coronary artery (RCA), and 1.84 +/- 0.23 cm for the left main artery (LM). The lengths statistics for Cartesian SSFP were 1.57 +/- 2.02 cm, 1.54 +/- 1.93 cm, 0.94 +/- 1.17 cm, 0.46 +/- 0.53 cm, respectively. The image sharpness was also increased from 0.61 +/- 0.13 (mm(-1)) in Cartesian-SSFP to 0.81 +/- 0.11 (mm(-1)) in VIPR-SSFP.

CONCLUSION: With VIPR trajectory the TR is substantially decreased, reducing the sensitivity of SSFP to field inhomogeneity and resulting in whole-heart images without banding artifacts at 3.0T. Image quality improved significantly over Cartesian sampling.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
20432361
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"