Report of the IFCC Working Group for Standardization of Thyroid Function Tests; part 1: thyroid-stimulating hormone.
Clinical Chemistry 2010 June
BACKGROUND: Laboratory testing of serum thyroid-stimulating hormone (TSH) is an essential tool for the diagnosis and management of various thyroid disorders whose collective prevalence lies between 4% and 8%. However, between-assay discrepancies in TSH results limit the application of clinical practice guidelines.
METHODS: We performed a method comparison study with 40 sera to assess the result comparability and performance attributes of 16 immunoassays.
RESULTS: Thirteen of 16 assays gave mean results within 10% of the overall mean. The difference between the most extreme means was 39%. Assay-specific biases could be eliminated by recalibration to the overall mean. After recalibration of singlicate results, all assays showed results within the biological total error goal (22.8%), except for 1 result in each of 4 assays. For a sample with a TSH concentration of 0.016 mIU/L, 6 assays either did not report results or demonstrated CVs >20%. Within-run and total imprecision ranged from 1.5% to 5.5% and 2.5% to 7.7%, respectively. Most assays were able to match the internal QC targets within 5%. Within-run drifts and shifts were observed.
CONCLUSIONS: Harmonization of TSH measurements would be particularly beneficial for 3 of the 16 examined assays. These data demonstrate that harmonization may be accomplished by establishing calibration traceability to the overall mean values for a panel of patient samples. However, the full impact of the approach must be further explored with a wider range of samples. Although a majority of assays showed excellent quality of performance, some would benefit from improved within-run stability.
METHODS: We performed a method comparison study with 40 sera to assess the result comparability and performance attributes of 16 immunoassays.
RESULTS: Thirteen of 16 assays gave mean results within 10% of the overall mean. The difference between the most extreme means was 39%. Assay-specific biases could be eliminated by recalibration to the overall mean. After recalibration of singlicate results, all assays showed results within the biological total error goal (22.8%), except for 1 result in each of 4 assays. For a sample with a TSH concentration of 0.016 mIU/L, 6 assays either did not report results or demonstrated CVs >20%. Within-run and total imprecision ranged from 1.5% to 5.5% and 2.5% to 7.7%, respectively. Most assays were able to match the internal QC targets within 5%. Within-run drifts and shifts were observed.
CONCLUSIONS: Harmonization of TSH measurements would be particularly beneficial for 3 of the 16 examined assays. These data demonstrate that harmonization may be accomplished by establishing calibration traceability to the overall mean values for a panel of patient samples. However, the full impact of the approach must be further explored with a wider range of samples. Although a majority of assays showed excellent quality of performance, some would benefit from improved within-run stability.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app