Stent graft versus balloon angioplasty for failing dialysis-access grafts.

BACKGROUND: The leading cause of failure of a prosthetic arteriovenous hemodialysis-access graft is venous anastomotic stenosis. Balloon angioplasty, the first-line therapy, has a tendency to lead to subsequent recoil and restenosis; however, no other therapies have yet proved to be more effective. This study was designed to compare conventional balloon angioplasty with an expanded polytetrafluoroethylene endovascular stent graft for revision of venous anastomotic stenosis in failing hemodialysis grafts.

METHODS: We conducted a prospective, multicenter trial, randomly assigning 190 patients who were undergoing hemodialysis and who had a venous anastomotic stenosis to undergo either balloon angioplasty alone or balloon angioplasty plus placement of the stent graft. Primary end points included patency of the treatment area and patency of the entire vascular access circuit.

RESULTS: At 6 months, the incidence of patency of the treatment area was significantly greater in the stent-graft group than in the balloon-angioplasty group (51% vs. 23%, P<0.001), as was the incidence of patency of the access circuit (38% vs. 20%, P=0.008). In addition, the incidence of freedom from subsequent interventions at 6 months was significantly greater in the stent-graft group than in the balloon-angioplasty group (32% vs. 16%, P=0.03 by the log-rank test and P=0.04 by the Wilcoxon rank-sum test). The incidence of binary restenosis at 6 months was greater in the balloon-angioplasty group than in the stent-graft group (78% vs. 28%, P<0.001). The incidences of adverse events at 6 months were equivalent in the two treatment groups, with the exception of restenosis, which occurred more frequently in the balloon-angioplasty group (P<0.001).

CONCLUSIONS: In this study, percutaneous revision of venous anastomotic stenosis in patients with a prosthetic hemodialysis graft was improved with the use of a stent graft, which appears to provide longer-term and superior patency and freedom from repeat interventions than standard balloon angioplasty. (ClinicalTrials.gov number, NCT00678249.)

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app