Evaluation in 3 months duration of neointimal coverage after zotarolimus-eluting stent implantation by optical coherence tomography: the ENDEAVOR OCT trial.

OBJECTIVES: We performed this study to investigate the vascular response in early period after zotarolimus-eluting stent (ZES) (Endeavor Sprint, Medtronic CardioVascular, Minneapolis, Minnesota) implantation.

BACKGROUND: The ZES has different characteristics, with biocompatible polymer and rapid drug-elution, compared with the first-generation drug-eluting stents (DES).

METHODS: The ENDEAVOR OCT (Evaluation in 3 Months Duration of Neointimal Coverage after Zotarolimus-Eluting Stent Implantation by Optical Coherence Tomography) trial is a prospective, single-center study evaluating vascular healing patterns with optical coherence tomography (OCT) at 3 months after stent implantation. A total of 31 ZES in 30 patients underwent serial OCT at immediate post-intervention and 3 months. Neointimal growth and malapposition were analyzed at each stent strut of cross-sectional OCT images with 0.5-mm intervals.

RESULTS: The incidence of malapposition at post-intervention and 3 months was 6.0% and 0.2%, respectively. However, late acquired malapposition was not detected at 3 months. Of 31 stents, 27 stents (87.1%) were covered completely with neointima, but the remaining 4 stents had 2 (0.8%), 4 (0.9%), 4 (1.2%), and 6 (1.4%) uncovered struts. Overall mean percentage of covered stent struts was 99.9 +/- 0.4%. This finding was consistent among groups with acute coronary syndrome and stable angina pectoris (99.9 +/- 0.3% vs. 99.9 +/- 0.4%, p = 0.92). Intracoronary thrombus was documented in 1 stent (3.2%) among 31 stents.

CONCLUSIONS: Most of the stent struts were covered with neointima, and late acquired malapposition was not found at 3 months after ZES implantation. Therefore, the current study demonstrated that ZES might have a favorable in vivo vascular response at 3 months after stent implantation. (Evaluation of Zotarolimus Eluting Stent at 3 Months Using Optical Coherence Tomography [ENDEAVOR OCT]; NCT00815139).

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app