STUDY DESIGN: Prospective case series
OBJECTIVE: This was designed to precisely measure motion after posterior dynamic stabilization using Dynesys instrumentation.
SUMMARY OF BACKGROUND DATA: The Dynesys posterior dynamic stabilization system, which stabilizes the spinal segment while potentially decreasing the risk of adjacent segment disease, is undergoing evaluation by the US Food and Drug Administration for treatment of degenerative spondylolisthesis without fusion. Evaluation of adjacent segment disease requires precise characterization of motion on the surgical level. Unfortunately, routine clinical radiographic techniques are imprecise and unreliable for full characterization of spinal segment motion. Radiostereometric analysis, which is very precise and reliable for in vivo measurement of motion, was used to examine spinal segment motion after dynamic stabilization with Dynesys.
METHODS: Six patients (age 59+/-7 y) underwent posterior decompression followed by posterior stabilization using Dynesys instrumentation (4 one-level, 2 two-levels). Three to 5 tantalum beads were placed in each vertebral body. Postoperative biplanar radiographs were obtained in flexion, extension, right, and left lateral bending, and 3-dimensional reconstruction was performed using radiostereometric analysis at 3, 6, 12, and 24 months postoperatively. The translations and rotations of the superior vertebral body were measured relative to the inferior vertebral body.
RESULTS: Over the 24-month follow-up period, mean flexion, extension, left, and right lateral bending of the motion segments were noted to be 1.0 degrees, 2.4 degrees, 0.6 degrees, and 0.6 degrees or less, respectively. There were no statistically significant changes in the degree of motion. During follow-up, no significant changes in neutral position of the device were noted in any of the 3 planes, and minimal translation was noted in the postoperative period.
CONCLUSIONS: The Dynesys dynamic instrumentation system seems to stabilize degenerative spondylolisthesis. As expected in the degenerative lumbar spine, the segmental motion of the implanted level in this study was limited and considerably less than normal spinal motion.
OBJECTIVE: This was designed to precisely measure motion after posterior dynamic stabilization using Dynesys instrumentation.
SUMMARY OF BACKGROUND DATA: The Dynesys posterior dynamic stabilization system, which stabilizes the spinal segment while potentially decreasing the risk of adjacent segment disease, is undergoing evaluation by the US Food and Drug Administration for treatment of degenerative spondylolisthesis without fusion. Evaluation of adjacent segment disease requires precise characterization of motion on the surgical level. Unfortunately, routine clinical radiographic techniques are imprecise and unreliable for full characterization of spinal segment motion. Radiostereometric analysis, which is very precise and reliable for in vivo measurement of motion, was used to examine spinal segment motion after dynamic stabilization with Dynesys.
METHODS: Six patients (age 59+/-7 y) underwent posterior decompression followed by posterior stabilization using Dynesys instrumentation (4 one-level, 2 two-levels). Three to 5 tantalum beads were placed in each vertebral body. Postoperative biplanar radiographs were obtained in flexion, extension, right, and left lateral bending, and 3-dimensional reconstruction was performed using radiostereometric analysis at 3, 6, 12, and 24 months postoperatively. The translations and rotations of the superior vertebral body were measured relative to the inferior vertebral body.
RESULTS: Over the 24-month follow-up period, mean flexion, extension, left, and right lateral bending of the motion segments were noted to be 1.0 degrees, 2.4 degrees, 0.6 degrees, and 0.6 degrees or less, respectively. There were no statistically significant changes in the degree of motion. During follow-up, no significant changes in neutral position of the device were noted in any of the 3 planes, and minimal translation was noted in the postoperative period.
CONCLUSIONS: The Dynesys dynamic instrumentation system seems to stabilize degenerative spondylolisthesis. As expected in the degenerative lumbar spine, the segmental motion of the implanted level in this study was limited and considerably less than normal spinal motion.
Full text links
Trending Papers
A Personalized Approach to the Management of Congestion in Acute Heart Failure.Heart International 2023
Potential Mechanisms of the Protective Effects of the Cardiometabolic Drugs Type-2 Sodium-Glucose Transporter Inhibitors and Glucagon-like Peptide-1 Receptor Agonists in Heart Failure.International Journal of Molecular Sciences 2024 Februrary 21
The Effect of Albumin Administration in Critically Ill Patients: A Retrospective Single-Center Analysis.Critical Care Medicine 2024 Februrary 8
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app