COMPARATIVE STUDY
JOURNAL ARTICLE

A novel noninvasive ultrasonic cardiac output monitor: comparison with cardiac magnetic resonance

Frederik Trinkmann, Christina Doesch, Theano Papavassiliu, Joerg Weissmann, Dariusch Haghi, Joachim Gruettner, Stefan O Schoenberg, Martin Borggrefe, Jens J Kaden, Joachim Saur
Clinical Cardiology 2010, 33 (2): E8-14
20043339

BACKGROUND: USCOM, a novel continuous wave Doppler (CWD) device, has been introduced for noninvasive determination of cardiac output (CO). The present study aimed to compare the accuracy and reproducibility of the new device, using cardiovascular magnetic resonance imaging (CMR) as the noninvasive gold standard.

METHODS AND RESULTS: The CO of 56 consecutive patients was prospectively determined by CWD either before or after CMR imaging. The CWD probe was placed in the suprasternal or supraclavicular notch aiming at the aortic valve. Valid CWD signals could be obtained in 45 patients yielding a CO of 5.3+/-1.1 L/min (range, 3.0-7.5 L/min) by CMR and 4.7+/-1.1 L/min by CWD (2.5-8.0 L/min, P = .004), respectively. CWD measurements showed an acceptable agreement with CMR (bias: 0.6+/-1.1 L/min) and a high reproducibility (bias: 0.1+/-0.4 L/min). Higher CO and body mass index (BMI) were identified as sources of inaccuracy in univariate analysis. By multivariate analysis, only CO(CMR) was found to be independently associated with larger variation. Estimated diameters of the left ventricular outflow tract (LVOT), a prerequisite for CO measurement by CWD, correlated only weakly with those measured by CMR.

CONCLUSIONS: Continuous wave Doppler is a feasible technique for measuring cardiac function. Although the overall agreement with CMR was acceptable, CWD showed a trend to underestimate CO. The estimated LVOT diameter by CWD is likely to be an important source of error. Nevertheless, the CWD device could be of clinical use especially for detection of intraindividual hemodynamic changes since a high reproducibility could be demonstrated.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
20043339
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"