Low-tube-voltage, high-tube-current multidetector abdominal CT: improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm--initial clinical experience.

Radiology 2010 January
PURPOSE: To investigate whether an adaptive statistical iterative reconstruction (ASIR) algorithm improves the image quality at low-tube-voltage (80-kVp), high-tube-current (675-mA) multidetector abdominal computed tomography (CT) during the late hepatic arterial phase.

MATERIALS AND METHODS: This prospective, single-center HIPAA-compliant study was institutional review board approved. Informed patient consent was obtained. Ten patients (six men, four women; mean age, 63 years; age range, 51-77 years) known or suspected to have hypervascular liver tumors underwent dual-energy 64-section multidetector CT. High- and low-tube-voltage CT images were acquired sequentially during the late hepatic arterial phase of contrast enhancement. Standard convolution FBP was used to reconstruct 140-kVp (protocol A) and 80-kVp (protocol B) image sets, and ASIR (protocol C) was used to reconstruct 80-kVp image sets. The mean image noise; contrast-to-noise ratio (CNR) relative to muscle for the aorta, liver, and pancreas; and effective dose with each protocol were assessed. A figure of merit (FOM) was computed to normalize the image noise and CNR for each protocol to effective dose. Repeated-measures analysis of variance with Bonferroni adjustment for multiple comparisons was used to compare differences in mean CNR, image noise, and corresponding FOM among the three protocols. The noise power spectra generated from a custom phantom with each protocol were also compared.

RESULTS: When image noise was normalized to effective dose, protocol C, as compared with protocols A (P = .0002) and B (P = .0001), yielded an approximately twofold reduction in noise. When the CNR was normalized to effective dose, protocol C yielded significantly higher CNRs for the aorta, liver, and pancreas than did protocol A (P = .0001 for all comparisons) and a significantly higher CNR for the liver than did protocol B (P = .003). Mean effective doses were 17.5 mSv +/- 0.6 (standard error) with protocol A and 5.1 mSv +/- 0.3 with protocols B and C. Compared with protocols A and B, protocol C yielded a small but quantifiable noise reduction across the entire spectrum of spatial frequencies.

CONCLUSION: Compared with standard FBP reconstruction, an ASIR algorithm improves image quality and has the potential to decrease radiation dose at low-tube-voltage, high-tube-current multidetector abdominal CT during the late hepatic arterial phase.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app