TASER conducted electrical weapons and implanted pacemakers and defibrillators.

INTRODUCTION: Conducted electrical weapons (CEW) have generated controversy in recent years regarding their effect on heart rhythm and on their suspected interaction with implanted devices such as the pacemakers and ICDs (implantable cardioverter defibrillators). We review the current evidence available on device interactions and pre-sent a new case series of 6 patients.

LITERATURE: We used the available case reports and animal studies on TASER or CEW related publications in PubMed.

CONCLUSION: Oversensing of TASER CEW discharges may cause noise reversion pacing in pacemakers and inappropriate detection of VF in ICDs. The nominal 5-second discharge is sufficiently short that neither clinically significant inhibition of bradycardia pacing nor inappropriate ICD shocks have been reported. Current evidence indicates that CEW discharges do not have adverse effects on pacemakers and ICDs.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app