Energy loss due to paravalvular leak with transcatheter aortic valve implantation

Ali N Azadani, Nicolas Jaussaud, Peter B Matthews, Liang Ge, T Sloane Guy, Timothy A M Chuter, Elaine E Tseng
Annals of Thoracic Surgery 2009, 88 (6): 1857-63
19932249

BACKGROUND: Mild to moderate paravalvular leaks commonly occur after transcatheter aortic valve (TAV) implantation. Current TAVs match and may exceed hemodynamic performance of surgically implanted bioprostheses based on pressure gradient and effective orifice area. However, these hemodynamic criteria do not account for paravalvular leaks. We recently demonstrated that TAV implantation within 23 mm Perimount bioprostheses (Edwards Lifesciences, Irvine, CA) yields similar hemodynamics to the 23 mm Perimount valve. However, mild paravalvular leakage was seen after TAV implantation. The present study quantifies energy loss during the entire cardiac cycle to assess the impact of TAV paravalvular leaks on the ventricle.

METHODS: Four TAVs designed to mimic the 23 mm SAPIEN valve (Edwards Lifesciences) were created. Transvalvular energy loss of 19, 21, and 23 mm Carpentier-Edwards bioprostheses were obtained in vitro within a pulse duplicator as a hemodynamic baseline (n = 4). The 23 mm TAVs were subsequently implanted within the 23 mm bioprostheses to assess energy loss due to paravalvular leak.

RESULTS: The 23 mm bioprosthesis demonstrated the least energy loss (213.25 +/- 31.35 mJ) compared with the 19 mm (330.00 +/- 36.97 mJ, p = 0.003) and 21 mm bioprostheses (298.00 +/- 37.25 mJ, p = 0.008). The TAV controls had similar energy loss (241.00 +/- 30.55 mJ, p = 0.17) as the 23 mm bioprostheses. However, after TAV implantation, total energy loss increased to 365.33 +/- 8.02 mJ significantly exceeding the energy loss of the 23 mm bioprosthesis (p < 0.001). Due to mild TAV paravalvular leakage, 39% of energy loss occurred during diastole.

CONCLUSIONS: Substantial energy loss during diastole occurs due to TAV paravalvular leakage. In the presence of mild paravalvular leakage, TAV implantation imposes a significantly higher workload on the left ventricle than an equivalently sized surgically implanted bioprosthesis.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
19932249
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"