70 vs 120 W thulium:yttrium-aluminium-garnet 2 microm continuous-wave laser for the treatment of benign prostatic hyperplasia: a systematic ex-vivo evaluation

Thorsten Bach, Nina Huck, Felix Wezel, Axel Häcker, Andreas J Gross, Maurice Stephan Michel
BJU International 2010, 106 (3): 368-72
19912204

OBJECTIVE: To evaluate the ablative and haemostatic properties of the recently introduced 120-W thulium:yttrium-aluminium-garnet (Tm-YAG) laser and to assess these results against those of the previously introduced 70-W Tm-YAG laser.

MATERIALS AND METHODS: The ex-vivo model of the isolated blood-perfused porcine kidney was used to determine the ablation capacity, haemostatic properties and coagulation depth of a 2 microm continuous-wave Tm-YAG laser. The energy was delivered using a 550-microm and an 800-microm bare-ended fibre. The results of the recently introduced 120-W Tm-YAG were compared to the established 70-W device. Kidney tissue was embedded for histological evaluation. After staining (haematoxylin and eosin, H & E; and NADH) of the specimen, the coagulation zone and depth of the necrotic tissue layer were measured.

RESULTS: With increased power output, the mean (sd) rate of vaporization of tissue increased, from 9.80 (3.03) g/10 min at 70 W to 16.41 (5.2) g/10 min at 120 W using the 550 microm fibre. The total amount of ablated tissue using the 800 microm fibre was lower than with the 550 microm fibre. With increasing power output the bleeding rate remained stable in either group. Tissue penetration remained shallow, even with increasing power output. In contrast to H&E staining, where the coagulation zone was measured, NADH staining showed an inner zone of necrotic tissue, again with no difference between the 70- and the 120-W Tm-YAG.

CONCLUSION: The 120-W Tm-YAG offers significantly higher ablation rates than the 70-W device, and despite the increased rate of ablation with the 120-W Tm-YAG, the bleeding rate and depth of tissue penetration were comparable to those using the 70-W device.

Comments

You need to log in or sign up for an account to be able to comment.

No comments yet, be the first to post one!

Related Papers

Available on the App Store

Available on the Play Store
Remove bar
Read by QxMD icon Read
19912204
×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"