Our purpose is to improve the performance sensitivity of a fiber sensor used as a fiber optic hydrophone probe (FOHP) by the addition of nanoscale thin film gold coating. The fiber is designed to provide a uniform and spatial averaging free response up to 100 MHz by etching down to an active diameter of approximately 9 mum. The performance sensitivity of straight cleaved (i.e., full size core and cladding) uncoated, tapered uncoated, and tapered thin film gold-coated fiber sensors was compared in the frequency range from 1.5 to 20 MHz in the presence of acoustic amplitude pressure levels as high as 6 MPa. An unprecedented voltage sensitivity of -245 dB relative to 1 V/muPa (560 mV/MPa) was measured for a thin film gold-coated FOHP by optimizing the gold coating thickness.