INTRODUCTION: Rapid intravascular access is a prerequisite component of emergency care and resuscitation. Peripheral intravenous (IV) access is the first-choice for most of the medical or trauma patients, but may be delayed in emergency conditions because of various difficulties. Elsewhere, intraosseous (IO) access may now be easily performed with a new semi-automatic battery-powered IO-insertion device (EZ-IO. The aim of this study was to compare the overall time to establish IO infusion with the EZ-IO device and the equivalent time for peripheral IV infusion, performed by emergency personnel in standard (No-CBRN) and in chemical, biological, radiological, and nuclear (CBRN) protective equipment.
METHODS: Nine nurses and 16 physicians randomly performed 4 procedures on a training manikin: IV and IO access under No-CBRN conditions and IV and IO under CBRN conditions. The time for each infusion attempt included all the steps essential for a simulated safe clinical use of infusion.
RESULTS: Under No-CBRN conditions, the time to establish IO infusion was shorter than the equivalent IV time (50+/-9 vs 70+/-30s). Similarly, under CBRN conditions, the time for IO infusion was shorter than for IV infusion (65+/-17 vs 104+/-30s). The mean time saved by IO infusion over IV infusion was respectively 20+/-24s (P<0.001) and 39+/-20s (P<0.001) under No-CBRN and CBRN conditions.
CONCLUSION: The time to establish IO infusion was significantly shorter than that for peripheral IV infusion, under both No-CBRN and CBRN conditions. Further clinical studies are required to confirm that IO access would effectively save time over IV access in real pre-hospital emergency settings.
METHODS: Nine nurses and 16 physicians randomly performed 4 procedures on a training manikin: IV and IO access under No-CBRN conditions and IV and IO under CBRN conditions. The time for each infusion attempt included all the steps essential for a simulated safe clinical use of infusion.
RESULTS: Under No-CBRN conditions, the time to establish IO infusion was shorter than the equivalent IV time (50+/-9 vs 70+/-30s). Similarly, under CBRN conditions, the time for IO infusion was shorter than for IV infusion (65+/-17 vs 104+/-30s). The mean time saved by IO infusion over IV infusion was respectively 20+/-24s (P<0.001) and 39+/-20s (P<0.001) under No-CBRN and CBRN conditions.
CONCLUSION: The time to establish IO infusion was significantly shorter than that for peripheral IV infusion, under both No-CBRN and CBRN conditions. Further clinical studies are required to confirm that IO access would effectively save time over IV access in real pre-hospital emergency settings.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app