Thoratec implantable ventricular assist device: the Papworth experience.

OBJECTIVE: The Thoratec (Thoratec Corp, Pleasanton, Calif) implantable ventricular assist device (IVAD) can be used for univentricular or biventricular support. The objective of this study is to review the 5-year experience of bridging patients to heart transplantation with this device in a single center. Surgical aspects, including hybrid pump pocket, double tunneling of driveline, and optimal cannulae placement, are discussed.

METHODS: This is a retrospective review of 24 patients treated between January 2002 and December 2007. Nineteen patients (79.1%) received a single implantable ventricular assist device as left ventricular assist devices, and 5 patients (21.9%) received 2 implantable ventricular assist devices as biventricular assist devices. The devices were implanted in pre-peritoneal/posterior rectus hybrid pump pockets. The driveline was passed through a 2-stage double-tunnel to exit onto the lateral chest wall. Patients were anticoagulated with Warfarin aiming for an international normalized ratio of 2.0 to 3.0.

RESULTS: Twenty male and 4 female patients with a mean age of 39.8 years (17-57 years) and a body surface area of 1.87 m(2) (1.63-2.2 m(2)) were supported for a total of 2308 patient-days. Mean duration of support was 96 days (10-301 days). The cause of heart failure was dilated cardiomyopathy in 18 patients and ischemic cardiomyopathy in 6 patients. Preoperatively, 23 patients were receiving inotropes, 12 patients required intra-aortic balloon pump support, 5 patients were intubated and mechanically ventilated, and 3 patients required continuous venovenous hemofiltration for renal support. Eleven patients (45.8%) were discharged with ventricular assist device support (1015 home patient-days). Complications observed were a) neurologic: stroke in 3 patients, transient ischemic attacks in 4 patients; and b) infection: driveline infection in 3 patients and pump pocket infection in 1 patient. There was no mechanical device failure. Support to transplantation was achieved in 17 patients (70.8%): 3 of 5 biventricular assist devices (60%) and 14 of 19 left ventricular assist devices (73.7%).

CONCLUSION: The Thoratec IVAD is a versatile and reliable ventricular assist device. It can provide univentricular or biventricular support for bridging patients to heart transplantation with acceptable complication rates. The portable Thoratec TLC-II console facilitated discharge while patients waited for a suitable donor heart.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app