Cyclic testing of 3 all-inside meniscal repair devices: a biomechanical analysis.

BACKGROUND: Despite the growing popularity of all-inside meniscal repair devices, concerns remain about their fixation strength. It is also unclear which of these devices have the most ideal biomechanical properties.

PURPOSE: To compare the biomechanical properties of 3 all-inside meniscal repair devices: the Meniscal Cinch, Ultra FAST-FIX, and MaxFire.

STUDY DESIGN: Controlled laboratory study.

METHODS: Twenty-seven human cadaveric menisci (3 groups of 9) were repaired using 3 different meniscal repair devices. The repaired menisci were then subjected to cyclic loading and load-to-failure testing. Gap formation and ultimate load to failure were measured.

RESULTS: Six of the devices failed during cyclic testing, 4 in the MaxFire group (44%), 1 in the Ultra FAST-FIX group (11%), and 1 in the Meniscal Cinch group (11%). After 1 cycle, there was a trend toward larger gap formation in the MaxFire group (3.65 mm) compared with the Meniscal Cinch group (2.12 mm, P = .05). After 100 cycles, group differences were found in gap formation (P = .03), with the MaxFire group exhibiting greater displacement (6.70 mm) than the Ultra FAST-FIX group (3.59 mm). After 500 cycles, group differences in gap formation (Meniscal Cinch, 5.94; Ultra FAST-FIX, 4.74 mm; Max Fire, 7.19 mm) did not reach statistical significance (P = .20). A trend was found toward higher ultimate load to failure in the Ultra FAST-FIX (86.1 N) and Meniscal Cinch (85.3 N) groups compared with the MaxFire group (64.5 N, P = .06). Stiffness was also higher in the Ultra FAST-FIX (25.2 N/mm) and Meniscal Cinch (25.5 N/mm) groups than the MaxFire group (16.3 N/mm, P = .02).

CONCLUSION: The Meniscal Cinch and Ultra FAST-FIX devices have more desirable biomechanical properties than the MaxFire as demonstrated by higher stiffness and a lower failure rate during cyclic testing.

CLINICAL RELEVANCE: The Meniscal Cinch and Ultra FAST-FIX devices may be more desirable implants for use during all-inside meniscal repair as they have superior biomechanical properties when compared with the MaxFire device.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app