We have located links that may give you full text access.
EVALUATION STUDIES
JOURNAL ARTICLE
Impact of skin tone on the performance of a transcutaneous jaundice meter.
Acta Paediatrica 2009 December
AIM: To evaluate the performance of the Konica Minolta/Air-Shields JM-103 jaundice meter on the basis of infant skin tone during the early neonatal period.
METHODS: Infants were prospectively categorized into light, medium and dark skin tone groups relative to two reference colours. Transcutaneous bilirubin readings were taken at predetermined intervals through the early neonatal period on a convenience sample of 938 healthy infants > or =37 weeks gestation. Serum bilirubin measurements were drawn routinely with metabolic studies and repeated in the presence of an elevated transcutaneous reading or clinically significant jaundice.
RESULTS: Multivariate linear regression analysis showed a significant impact on serum and transcutaneous bilirubin agreement by skin tone. Highest precision and lowest bias were observed for medium skin toned infants. Greater disagreement between serum and transcutaneous measurements was noted at serum bilirubin concentrations >200 micromol/L. Insufficient numbers of dark skin toned infants were enrolled to evaluate fully the performance of the jaundice meter for this group.
CONCLUSION: The JM-103 jaundice meter displayed good correlation with serum bilirubin concentrations in light and medium skin tone infants, although it showed a tendency to under-read in the lighter skin tone group and to over-read in the darker skin tone group. The device shows excellent performance characteristics for use as a screening device.
METHODS: Infants were prospectively categorized into light, medium and dark skin tone groups relative to two reference colours. Transcutaneous bilirubin readings were taken at predetermined intervals through the early neonatal period on a convenience sample of 938 healthy infants > or =37 weeks gestation. Serum bilirubin measurements were drawn routinely with metabolic studies and repeated in the presence of an elevated transcutaneous reading or clinically significant jaundice.
RESULTS: Multivariate linear regression analysis showed a significant impact on serum and transcutaneous bilirubin agreement by skin tone. Highest precision and lowest bias were observed for medium skin toned infants. Greater disagreement between serum and transcutaneous measurements was noted at serum bilirubin concentrations >200 micromol/L. Insufficient numbers of dark skin toned infants were enrolled to evaluate fully the performance of the jaundice meter for this group.
CONCLUSION: The JM-103 jaundice meter displayed good correlation with serum bilirubin concentrations in light and medium skin tone infants, although it showed a tendency to under-read in the lighter skin tone group and to over-read in the darker skin tone group. The device shows excellent performance characteristics for use as a screening device.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app