The effect of activity outside the field of view on image quality for a 3D LSO-based whole body PET/CT scanner.

The purpose of this study was to quantify the influence of outside field of view (FOV) activity concentration (A(c)(,out)) on the noise equivalent count rate (NECR), scatter fraction (SF) and image quality of a 3D LSO whole-body PET/CT scanner. The contrast-to-noise ratio (CNR) was the figure of merit used to characterize the image quality of PET scans. A modified International Electrotechnical Commission (IEC) phantom was used to obtain SF and counting rates similar to those found in average patients. A scatter phantom was positioned at the end of the modified IEC phantom to simulate an activity that extends beyond the scanner. The modified IEC phantom was filled with (18)F (11 kBq mL(-1)) and the spherical targets, with internal diameter (ID) ranging from 10 to 37 mm, had a target-to-background ratio of 10. PET images were acquired with background activity concentrations into the FOV (A(c)(,bkg)) about 11, 9.2, 6.6, 5.2 and 3.5 kBq mL(-1). The emission scan duration (ESD) was set to 1, 2, 3 and 4 min. The tube inside the scatter phantom was filled with activities to provide A(c)(,out) in the whole scatter phantom of zero, half, unity, twofold and fourfold the one of the modified IEC phantom. Plots of CNR versus the various parameters are provided. Multiple linear regression was employed to study the effects of A(c)(,out) on CNR, adjusted for the presence of variables (sphere ID, A(c)(,bkg) and ESD) related to CNR. The presence of outside FOV activity at the same concentration as the one inside the FOV reduces peak NECR of 30%. The increase in SF is marginal (1.2%). CNR diminishes significantly with increasing outside FOV activity, in the range explored. ESD and A(c)(,out) have a similar weight in accounting for CNR variance. Thus, an experimental law that adjusts the scan duration to the outside FOV activity can be devised. Recovery of CNR loss due to an elevated A(c)(,out) activity seems feasible by modulating the ESD in individual bed positions according to A(c)(,out).

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app