MRI guidance for accelerated partial breast irradiation in prone position: imaging protocol design and evaluation.
International Journal of Radiation Oncology, Biology, Physics 2009 September 2
PURPOSE: To design and evaluate a magnetic resonance imaging (MRI) protocol to be incorporated in the simulation process for external beam accelerated partial breast irradiation.
METHODS AND MATERIALS: An imaging protocol was developed based on an existing breast MRI technique with the patient in the prone position on a dedicated coil. Pulse sequences were customized to exploit T1 and T2 contrast mechanisms characteristic of lumpectomy cavities. A three-dimensional image warping algorithm was included to correct for geometric distortions related to nonlinearity of spatially encoding gradients. Respiratory motion, image distortions, and susceptibility artifacts of 3.5-mm titanium surgical clips were examined. Magnetic resonance images of volunteers were acquired repeatedly to analyze residual setup deviations resulting from breast tissue deformation.
RESULTS: The customized sequences generated high-resolution magnetic resonance images emphasizing lumpectomy cavity morphology. Respiratory motion was negligible with the subject in the prone position. The gradient-induced nonlinearity was reduced to less than 1 mm in a region 15 cm away from the isocenter of the magnet. Signal-void regions of surgical clips were 4 mm and 8 mm for spin echo and gradient echo images, respectively. Typical residual repositioning errors resulting from breast deformation were estimated to be 3 mm or less.
CONCLUSIONS: MRI guidance for accelerated partial breast irradiation with the patient in the prone position with adequate contrast, spatial fidelity, and resolution is possible.
METHODS AND MATERIALS: An imaging protocol was developed based on an existing breast MRI technique with the patient in the prone position on a dedicated coil. Pulse sequences were customized to exploit T1 and T2 contrast mechanisms characteristic of lumpectomy cavities. A three-dimensional image warping algorithm was included to correct for geometric distortions related to nonlinearity of spatially encoding gradients. Respiratory motion, image distortions, and susceptibility artifacts of 3.5-mm titanium surgical clips were examined. Magnetic resonance images of volunteers were acquired repeatedly to analyze residual setup deviations resulting from breast tissue deformation.
RESULTS: The customized sequences generated high-resolution magnetic resonance images emphasizing lumpectomy cavity morphology. Respiratory motion was negligible with the subject in the prone position. The gradient-induced nonlinearity was reduced to less than 1 mm in a region 15 cm away from the isocenter of the magnet. Signal-void regions of surgical clips were 4 mm and 8 mm for spin echo and gradient echo images, respectively. Typical residual repositioning errors resulting from breast deformation were estimated to be 3 mm or less.
CONCLUSIONS: MRI guidance for accelerated partial breast irradiation with the patient in the prone position with adequate contrast, spatial fidelity, and resolution is possible.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app