A comparison of stroke volume variation measured by Vigileo/FloTrac system and aortic Doppler echocardiography.
Anesthesia and Analgesia 2009 August
BACKGROUND: The goal of this study was to compare stroke volume variation (SVV) assessed from a peripheral artery with the Vigileo/FloTrac system (SVV-FloTrac) with SVV derived close to the heart by aortic Doppler (SVV-Doppler).
METHODS: Thirty patients undergoing liver transplantation underwent simultaneous SVV-FloTrac and SVV-Doppler measurements before and after intravascular volume expansion.
RESULTS: SVV-FloTrac and SVV-Doppler comparison before intravascular volume expansion showed a mean bias of 0.7%, and 95% limits of agreement of -4.2% to 5.5%. The areas under the receiver operating characteristic curves generated to discriminate responders and nonresponders to intravascular volume expansion were not different for SVV-FloTrac and SVV-Doppler.
CONCLUSIONS: SVV-FloTrac and SVV-Doppler measurements show acceptable bias and limits of agreement, and similar performance in terms of fluid responsiveness in patients undergoing liver transplantation.
METHODS: Thirty patients undergoing liver transplantation underwent simultaneous SVV-FloTrac and SVV-Doppler measurements before and after intravascular volume expansion.
RESULTS: SVV-FloTrac and SVV-Doppler comparison before intravascular volume expansion showed a mean bias of 0.7%, and 95% limits of agreement of -4.2% to 5.5%. The areas under the receiver operating characteristic curves generated to discriminate responders and nonresponders to intravascular volume expansion were not different for SVV-FloTrac and SVV-Doppler.
CONCLUSIONS: SVV-FloTrac and SVV-Doppler measurements show acceptable bias and limits of agreement, and similar performance in terms of fluid responsiveness in patients undergoing liver transplantation.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app