Assessment of artifacts and reproducibility across spectral- and time-domain optical coherence tomography devices.

Ophthalmology 2009 October
PURPOSE: To report the frequency of optical coherence tomography (OCT) scan artifacts and to compare macular thickness measurements, interscan reproducibility, and interdevice agreeability across 3 spectral-domain (SD) OCT (also known as Fourier domain; Cirrus HD-OCT, RTVue-100, and Topcon 3D-OCT 1000) devices and 1 time-domain (TD) OCT (Stratus OCT) device.

DESIGN: Prospective, noncomparative, noninterventional case series.

PARTICIPANTS: Fifty-two patients seen at the New England Eye Center, Tufts Medical Center Retina Service, between February and August 2008.

METHODS: Two scans were performed for each of the SD OCT protocols: Cirrus macular cube 512 x 128 (software version 3.0; Carl Zeiss Meditec, Inc., Dublin, CA), RTVue (E)MM5 and MM6 (software version 3.5; Optovue, Inc., Fremont, CA), Topcon 3D Macular and Radial (software version 2.12; Topcon, Inc., Paramus, NJ), in addition to 1 TD OCT scan via Stratus macular thickness protocol (software version 4.0; Carl Zeiss Meditec, Inc.). Scans were inspected for 6 types of OCT scan artifacts and were analyzed. Interscan reproducibility and interdevice agreeability were assessed by intraclass correlation coefficients (ICCs) and Bland-Altman plots, respectively.

MAIN OUTCOME MEASURES: Optical coherence tomography image artifacts, macular thickness, reproducibility, and agreeability.

RESULTS: Time-domain OCT scans contained a significantly higher percentage of clinically significant improper central foveal thickness (IFT) after manual correction (11-mum change or more) compared with SD OCT scans. Cirrus HD-OCT had a significantly lower percentage of clinically significant IFT (11.1%) compared with the other SD OCT devices (Topcon 3D, 20.4%; Topcon Radial, 29.6%; RTVue (E)MM5, 42.6%; RTVue MM6, 24.1%; P = 0.001). All 3 SD OCT devices had central foveal subfield thicknesses that were significantly more than that of TD OCT after manual correction (P<0.0001). All 3 SD OCT devices demonstrated a high degree of reproducibility in the central foveal region (ICCs, 0.92-0.97). Bland-Altman plots showed low agreeability between TD and SD OCT scans.

CONCLUSIONS: Out of all OCT devices analyzed, cirrus HD-OCT scans exhibited the lowest occurrence of any artifacts (68.5%), IFT (40.7%), and clinically significant IFT (11.1%), whereas Stratus OCT scans exhibited the highest occurrence of clinically significant IFT. Further work on improving segmentation algorithm to decrease artifacts is warranted.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app