Comparison of intraocular lens power calculation by the IOLMaster in phakic and eyes with hydrophobic acrylic lenses.

Ophthalmology 2009 July
PURPOSE: To compare the optical biometry measurements and intraocular lens (IOL) power estimation using the IOLMaster (Carl Zeiss Meditec, Dublin, CA) in phakic eyes and eyes with hydrophobic acrylic lenses.

DESIGN: Observational cross-sectional study.

PARTICIPANTS: A total of 156 patients (226 eyes).

METHODS: The IOLMaster measurements (IOLM-1) were performed before phacoemulsification and reexamined 3 months postoperatively (IOLM-2). One of the foldable acrylic IOLs (AcrySof SA60AT, SN60WF, or SN60D3, Alcon Laboratories Inc., Dallas, TX) was implanted.

MAIN OUTCOME MEASURES: The expected refraction and estimation error calculated from IOLM-1 using the SRK II, SRK/T, and Haigis formulae were compared with the residual refraction 3 months postoperatively. The power of the implanted IOL and IOLM-2 measurement data were used to re-estimate the postoperative expected refraction in pseudophakic eyes. The difference in expected refraction and estimation error between phakic and pseudophakic eyes was studied. Differences in the anterior chamber depth and axial length measured by IOLM-1 and IOLM-2 were analyzed and correlated with the estimation error.

RESULTS: The IOLMaster measured an axial length 0.10+/-0.15 mm shorter in pseudophakic eyes (P<0.001). Calculations from IOLM-2 gave a significantly more hyperopic expected refraction than IOLM-1, with an averaged 0.20+/-0.46 diopters (D), 0.18+/-0.45 D, and 0.65+/-0.59 D calculated by the SRK II, SRK/T, and Haigis formulae, respectively. There was no significant difference among the 3 IOLs. The difference in estimation error correlated with the difference in axial length and anterior chamber depth (P<0.001 for the SRK II, SRK/T, and Haigis formulae). However, the correlation was strongest when the Haigis formula was used for the calculation.

CONCLUSIONS: The expected refraction in pseudophakic eyes differed significantly from that in phakic eyes by the IOLMaster depending on the IOL formulae used for the calculation rather than the type of IOL. An adjustment of target refraction by 0.20 to 0.65 D toward the hyperopic side of the desired refraction could be considered when using optical biometry data in pseudophakic eyes to achieve postoperative emmetropia.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app