JOURNAL ARTICLE

Evaluation of the LMA Supreme in 100 non-paralysed patients

T M Cook, J J Gatward, J Handel, R Hardy, C Thompson, R Srivastava, P A Clarke
Anaesthesia 2009, 64 (5): 555-62
19413827
We studied the LMA Supreme in 100 elective, anaesthetised, healthy patients assessing: ease of use, airway quality, anatomical and functional positioning, airway leak and complications. Insertion was successful on first, second or third attempt in 90, nine and one patient respectively. Thirty manipulations were required in 22 patients to achieve a clear airway. Median [interquartile (range)] insertion time was 18 [10-25 (5-120)] s. During ventilation, an expired tidal volume of 7 ml x kg(-1) was achieved in all patients. Median [interquartile (range)] airway leak pressure was 24 [20-28 (13-40)] cmH(2)O. On fibreoptic examination via the device, vocal cords were visible in 83 patients (85%). During maintenance, five patients (5%) required 13 airway manipulations. There was one episode of minor regurgitation, without aspiration. Other complications and patient side-effects were mild and few. The LMA Supreme is easily and rapidly inserted, providing a reliable airway and good airway seal. Further studies are indicated to assess safety and performance compared to other supraglottic airway devices.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
19413827
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"