The QuantumCor device for treating mitral regurgitation: an animal study

Ramil Goel, Thomas Witzel, Duane Dickens, Patricia A Takeda, Richard R Heuser
Catheterization and Cardiovascular Interventions 2009 July 1, 74 (1): 43-8
19412916

OBJECTIVES: The mitral annular contraction achieved could help reduce mitral regurgitation (MR), and with appropriate modifications, be applied to human subjects providing a potentially effective percutaneous method of valve repair.

BACKGROUND: MR is an important source of morbidity and is an independent predictor of mortality. A variety of percutaneous approaches are being developed to address this issue. We introduce a novel potential method utilizing radiofrequency (RF) energy to heat and shrink the mitral valve annulus in an animal model.

METHODS: In open-heart procedures in 16 healthy sheep (six with naturally occurring MR), we used a malleable probe (QuantumCor, Lake Forest, CA) that conforms to the annular shape to deliver RF energy via a standard generator to replicate a surgical mitral annular ring. Seven sheep were followed chronically and their mitral annulus dimensions measured serially.

RESULTS: All sheep underwent intracardiac echocardiography or direct circumferential measurement of the mitral annulus before and after RF therapy. RF therapy was administered in less than 4 min in each case, and the mean anteroposterior (AP) annular distance was reduced by a mean of 23.8% (AP diameter reduction 5.75 +/- 0.86 mm, P < 0.001) acutely. In the six sheep with nonischemic MR, regurgitation was eliminated. Acute histopathology (HP) demonstrated no damage to the leaflets, coronary sinuses, or coronary arteries. At the end of the intended 6-month period of the chronic part of the study, four of the seven animals survived. The four treatment animals showed significant reductions in mitral A-P dimension, with a percent diameter reduction of 26.4% (AP diameter reduction 7 +/- 2.3 mm).

CONCLUSION: The application of RF directly to heat the mitral annulus has resulted in sustained contraction of the annulus in this limited preclinical animal study. With further study and possible modifications, it holds promise for future application in human subjects with MR.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
19412916
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"

We want to hear from doctors like you!

Take a second to answer a survey question.