JOURNAL ARTICLE
RESEARCH SUPPORT, NON-U.S. GOV'T
Add like
Add dislike
Add to saved papers

The adjustable continence therapy system for recurrent female stress urinary incontinence: 1-year results of the North America Clinical Study Group.

PURPOSE: We determined the efficacy, safety, adjustability and technical feasibility of the adjustable continence therapy device (Uromedica, Plymouth, Minnesota) for the treatment of recurrent female stress urinary incontinence.

MATERIALS AND METHODS: Female patients with recurrent stress urinary incontinence were enrolled in the study and a defined set of exclusionary criteria were followed. Baseline and regular followup tests to determine eligibility, and to measure subjective and objective improvement were performed. A trocar was passed fluoroscopically and with digital vaginal guidance to the urethrovesical junction through small incisions between the labia majora and minora. The adjustable continence therapy device was delivered and the balloons were filled with isotonic contrast. The injection ports for balloon inflation were placed in a subcutaneous pocket in each labia majora. Device adjustments were performed percutaneously in the clinic postoperatively. An approved investigational device exemption Food and Drug Administration protocol was followed to record all adverse events.

RESULTS: A total of 162 subjects underwent implantation with 1 year of data available on 140. Mean Stamey score improved by 1 grade or more in 76.4% (107 of 140) of subjects. Improvement in the mean incontinence quality of life questionnaire score was noted at 36.5 to 70.7 (p <0.001). Reductions in mean Urogenital Distress Inventory (60.3 to 33.4) and Incontinence Impact Questionnaire (54.4 to 23.4) scores also occurred (p <0.001). Mean provocative pad weight decreased from 49.6 to 11.2 gm (p <0.001). Of the patients 52% (67 of 130) were dry at 1 year (less than 2 gm on provocative pad weight testing) and 80% (102 of 126) were improved (greater than 50% reduction on provocative pad weight testing). Complications occurred in 24.4% (38 of 156) of patients. Explantation was required in 18.3% (28 of 153) of the patients during 1 year. In terms of the complications 96.0% were considered to be mild or moderate.

CONCLUSIONS: The Uromedica adjustable continence therapy device is an effective, simple, safe and minimally invasive treatment for recurrent female stress urinary incontinence. It can be easily adjusted percutaneously to enhance efficacy and complications are usually easily manageable. Explantation does not preclude later repeat implantation.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app