Assessment of the absorption process following bioabsorbable everolimus-eluting stent implantation: temporal changes in strain values and tissue composition using intravascular ultrasound radiofrequency data analysis. A substudy of the ABSORB clinical trial.

EuroIntervention 2009 January
AIMS: The main objective was to use IVUS-backscatter radiofrequency (IVUS-RF) to assess the degradation of a bioabsorbable stent by measuring serial changes in dense calcium (DC) and necrotic core (NC) as assessed by intravascular ultrasound-Virtual Histology (IVUS-VH) and in the strain as assessed by palpography.

METHODS AND RESULTS: In the ABSORB trial, 27 patients treated with a single bioabsorbable everolimus-eluting stent (BVS, Abbott Vascular, Santa Clara, CA, USA) were all imaged with IVUS-RF post-stenting and at 6-month follow-up, and 13 and 12 patients were also investigated pre-stenting with IVUS-VH and palpography respectively. From pre- to post-stenting, with VH (n = 13), there was an increase in mean "DC" (9.8 vs. 25.4%, p = 0.0002) and "NC" (15.5 vs. 30.5%, p = 0.0002). In palpography (n = 12), the mean number of frames with Rotterdam Classification (ROC) III/IV per cm decreased from 1.22 +/- 1.91 to 0.12 +/- 0.31 (p = 0.0781) and the mean cumulative strain values (all frames with ROC I-IV scores) changed from 0.50 +/- 0.27 to 0.20 +/- 0.10% (p = 0.0034). Comparing post-stenting with follow-up (n = 27), VH showed a decrease in "DC" (29.7% vs. 21.1%, p = 0.0001). "NC" also decreased (26.9 vs. 21.5%, p = 0.0027). For palpography (n = 25 patients), an increase in the mean number of frames with ROC III/IV per cm was observed from 0.09 +/- 0.26 to 0.22 +/- 0.36 (p = -0.1563) while the mean cumulative strain values (all frames with ROC I-IV scores) changed from 0.15 +/- 0.10 to 0.26 +/- 0.12% (p < 0.0001).

CONCLUSIONS: IVUS-VH changes at 6 months suggest alteration of the BVS with reduction of RF backscattering by polymeric struts. Strained plaques on the palpograms were almost abolished following stent implantation. However, strain values reappeared within 6 months suggesting an increase in endoluminal deformability of the stented vessel.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app