Effects of endotracheal tube size and ventilator settings on the mechanics of a test system during intermittent flow ventilation.
The effect of varying the size of standard neonatal endotracheal tubes on delivered tidal volumes (VT), resistance (R), dynamic compliance (Cdyn), and resistive work of breathing (WOB) was measured in a test system during intermittent flow ventilation at different ventilator settings. The experiments were performed with a Sechrist infant ventilator connected to a Dräger Test Lung via standard neonatal endotracheal tubes. R, inspiratory (Ri), and expiratory resistance (Re) as well as WOB were significantly affected by endotracheal tube size. The calculated difference in Re between endotracheal tubes of 2.5 and 3.5 mm I.D. was 93 cm H2O/L/S (mean value for all studies). Cdyn and VT were also affected by endotracheal tube size. However, although statistically significant differences were found in Cdyn (mean, 0.584 mL/cm H2O and 0.567 mL/cm H2O) and VT (mean, 13.0 mL and 12.7 mL) for the tube sizes 3.5 mm and 2.5 mm I.D., respectively, the absolute numerical differences were small. Also, ventilator settings with respect to the peak inspiratory pressure (PIP) - positive end-expiratory pressure (PEEP) difference had a significant influence on Cdyn for both tube sizes. On the other hand, flow and inspiratory time adjustments had no significant effect on ventilatory parameters. Endotracheal tube size and ventilator settings should be considered when comparing the pulmonary function tests in intubated and non-intubated newborn infants.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app