Automated emergency ventilation devices in a simulated unprotected airway

Holger Herff, Christian A Schmittinger, Achim von Goedecke, Peter Paal, Thomas Mitterlechner, Karl H Lindner, Volker Wenzel
Journal of Emergency Medicine 2011, 41 (3): 246-51
19201138

BACKGROUND: Automated ventilation devices are becoming more popular for emergency ventilation, but there is still not much experience concerning the optimal ventilation mode.

METHODS: In a bench model representing a non-intubated patient in respiratory and cardiac arrest, we compared a pressure-cycled with a time- and volume-cycled automated ventilation device in their completely automated modes. The main study endpoints were inspiratory time, respiratory rate, stomach inflation, and lung tidal volumes.

RESULTS: The pressure-cycled device inspired for 6.7 s in the respiratory arrest setting (respiratory rate 5.6/min), and never reached its closing pressure in the cardiac arrest setting (respiratory rate 1 breath/min). The time- and volume-cycled device inspired in both settings for 1.7 s (respiratory rate 13 breaths/min). In the respiratory arrest setting, mask leakage was 620 ± 20 mL for the pressure-cycled device vs. 290 ± 10 mL for the time- and volume-cycled device (p < 0.0001); lung tidal volume was 1080 ± 50 mL vs. 490 ± 20 mL, respectively (p < 0.0001); and there was no stomach inflation for either device. In the cardiac arrest setting, pressure-cycled device mask leakage was 5460 ± 60 mL vs. 240 ± 20 mL (p < 0.0001) for the time- and volume-cycled device (p < 0.0001); stomach inflation was 13,100 ± 100 mL vs. 90 ± 10 mL, respectively (p < 0.0001); and lung tidal volume 740 ± 60 mL vs. 420 ± 20 mL, respectively (p < 0.0001).

CONCLUSION: In a simulated respiratory arrest setting, ventilation with an automated pressure-cycled ventilation device resulted in lower respiratory frequency and larger tidal volumes compared to a time- and volume-cycled device. In a simulated cardiac arrest setting, ventilation with an automated pressure-cycled ventilation device, but not a time- and volume-cycled device, resulted in continuous gastric insufflation.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
19201138
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"