Determinants of lesion dimensions during transcatheter microwave ablation.

BACKGROUND: Transcatheter microwave ablation is a novel technique for treating cardiac arrhythmias.

METHODS: We investigated the effects of catheter temperature, application duration, and antenna length on lesion dimensions during catheter-based microwave ablation. In a swine thigh muscle preparation, microwave was delivered at targeted temperatures of 60 degrees C (n = 18), 70 degrees C (n = 27), 80 degrees C (n = 43), or 90 degrees C (n = 18) for 120 seconds with 10-mm antenna; and at targeted temperatures of 80 degrees C for 120 seconds (n = 22), 150 seconds (n = 18), 180 seconds (n = 18), 210 seconds (n = 18), and 240 seconds (n = 17) with 20-mm antenna using 10 F catheter (MedWaves, San Diego, CA, USA) during parallel orientation. Conventional radiofrequency ablation (RF) using a 4-mm tip electrode was performed as control.

RESULTS: With 120-second energy applications, lesion length and depth were significantly larger with targeted temperatures of 80 degrees C and 90 degrees C than 60 degrees C (P< 0.05). Furthermore, lesion depth and width, but not length, were significantly increased by prolonging energy application duration from 120 to 240 seconds at targeted temperature of 80 degrees C (P< 0.05). Compared to RF, microwave lesions were significantly longer but had comparable depth and width. A 20-mm microwave antenna produced longer lesions than either a 10-mm antenna or RF ablation catheter. Multivariate analysis demonstrated that targeted temperature >or=80 degrees C, application duration >or=150 seconds, and use of 20-mm antenna were independent predictors for lesion depth and width (P< 0.05). Surface dessication was observed in 4/18 (22%) lesions at 90 degrees C, as compared with 1/136 (0.7%) at 80 degrees C targeted tip temperature (P < 0.05).

CONCLUSIONS: This study demonstrated that lesions size with transcatheter microwave ablation can be controlled by adjusting targeted temperature, energy application duration, and antenna length. A targeted temperature of 80 degrees C for more than 150 seconds should provide optimal lesion dimensions and lower risk of surface dessication or charring.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app