Assessment of changes in distribution of lung perfusion by electrical impedance tomography

Inéz Frerichs, Sven Pulletz, Gunnar Elke, Florian Reifferscheid, Dirk Schadler, Jens Scholz, Norbert Weiler
Respiration; International Review of Thoracic Diseases 2009, 77 (3): 282-91
19147986

BACKGROUND: Electrical impedance tomography (EIT) is able to detect variations in regional lung electrical impedance associated with changes in both air and blood content and potentially capable of assessing regional ventilation-perfusion relationships. However, regional lung perfusion is difficult to determine because the impedance changes synchronous with the heart rate are of very small amplitude.

OBJECTIVES: The aim of our study was to determine the redistribution of lung perfusion elicited by one-lung ventilation using EIT with a novel region-of-interest analysis.

METHODS: Ten patients (65 +/- 9 years, mean age +/- SD) scheduled for elective chest surgery were studied after intubation with a double-lumen endotracheal tube during bilateral and unilateral ventilation of the right and left lungs. EIT data were acquired at a rate of 25 scans/s. Relative impedance changes synchronous with the heart rate were evaluated in the right and left lung regions.

RESULTS: During bilateral ventilation, the mean right-to-left lung ratio of the sum of heart rate-related impedance changes was 1.12 +/- 0.20, but the ratio significantly changed (0.81 +/- 0.16 and 1.48 +/- 0.37) during unilateral left- and right-lung ventilation with reduced perfusion of the non-ventilated lung. Increased perfusion most likely occurred in the ventilated lung because the impedance values summed over both regions did not change (0.62 +/- 0.23 vs. 0.58 +/- 0.22) compared with bilateral ventilation.

CONCLUSIONS: Our results indicate that redistribution of regional lung perfusion can be assessed by EIT during one-lung ventilation. The performance of EIT in detecting changes in lung perfusion in even smaller lung regions remains to be established.

Full Text Links

Find Full Text Links for this Article

Discussion

You are not logged in. Sign Up or Log In to join the discussion.

Related Papers

Remove bar
Read by QxMD icon Read
19147986
×

Save your favorite articles in one place with a free QxMD account.

×

Search Tips

Use Boolean operators: AND/OR

diabetic AND foot
diabetes OR diabetic

Exclude a word using the 'minus' sign

Virchow -triad

Use Parentheses

water AND (cup OR glass)

Add an asterisk (*) at end of a word to include word stems

Neuro* will search for Neurology, Neuroscientist, Neurological, and so on

Use quotes to search for an exact phrase

"primary prevention of cancer"
(heart or cardiac or cardio*) AND arrest -"American Heart Association"